Long-term in vivo left ventricular assist device study for 284 days with Gyro PI pump

被引:13
作者
Ohtsuka, G
Nakata, K
Yoshikawa, M
Takano, T
Glueck, J
Sankai, Y
Takami, Y
Mueller, J
Sueoka, A
Letsou, G
Schima, H
Schmallegger, H
Wolner, E
Koyanagi, H
Fujisawa, A
Baldwin, JC
Nosé, Y
机构
[1] Baylor Coll Med, Dept Surg, Houston, TX 77030 USA
[2] Univ Vienna, Vienna, Austria
[3] Tokyo Womens Med Coll, Tokyo 162, Japan
[4] Kyocera Corp, Kyoto, Japan
关键词
implantable artificial heart; centrifugal heart;
D O I
10.1046/j.1525-1594.1999.06391.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A totally implantable centrifugal artificial heart has been developed. The plastic prototype, the Gyro PI 601, passed 2 day hemodynamic tests as a functional total artificial heart (TAH), 2 week screening tests for anti-thrombogenecity, and a 1 month system feasibility study. Based upon these results, a metallic prototype, the Gyro PI700 series, was subjected to long-term in vivo left ventricular assist device (LVAD) studies of over 1 month. The Gyro PI700 series has the same inner dimension and same characteristics of the Gyro PI 601 such as an eccentric inlet port, a double pivot bearing system, and a magnet coupling system. The PI metallic pump is also driven with the Vienna DC brushless motor actuator like the PI 601. The pump-actuator package was implanted in 3 calves in the preperitoneal space, bypassing from the left ventricular (LV) apex to the descending aorta. Case 1 achieved a 284 day survival. Case 2 was euthanized early at 72 postoperative days as a result of the functional obstruction of the inlet port due to the excessive growth of the calf. There was no blood clot inside the pumps of either case. Case 3 is on-going (22 days on July 24, 1998). During these periods, all cases showed no physiological abnormalities. In conclusion, the PI700 series pump has excellent results as a long-term implantable LVAD.
引用
收藏
页码:504 / 507
页数:4
相关论文
共 15 条
[1]  
FUJISAWA A, 1998, SEITAIZAIRYOU, V16, P1
[2]   Evaluation of the wear of the pivot bearing in the Gyro C1E3 pump [J].
Makinouchi, K ;
Nakazawa, T ;
Takami, Y ;
Takatani, S ;
Nose, Y .
ARTIFICIAL ORGANS, 1996, 20 (06) :523-528
[3]   Development and initial testing of a permanently implantable centrifugal pump [J].
Nakazawa, T ;
Takami, Y ;
Benkowski, R ;
Ohtsubo, S ;
Yukio, O ;
Tayama, E ;
Ohtsuka, G ;
Niimi, Y ;
Glueck, J ;
Sueoka, A ;
Schmallegger, H ;
Schima, H ;
Wolner, E ;
Nose, Y .
ARTIFICIAL ORGANS, 1997, 21 (07) :597-601
[4]  
Nojiri C, 1997, ASAIO J, V43, pM548
[5]   Therapeutic and physiological artificial heart: Future prospects [J].
Nose, Y ;
Ohtsubo, S ;
Tayama, E .
ARTIFICIAL ORGANS, 1997, 21 (07) :592-596
[6]  
NOSE Y, IN PRESS ANN THORAC
[7]   AN ULTIMATE, COMPACT, SEALLESS CENTRIFUGAL VENTRICULAR ASSIST DEVICE - BAYLOR C-GYRO PUMP [J].
OHARA, Y ;
MAKINOUCHI, K ;
ORIME, Y ;
TASAI, K ;
NAITO, K ;
MIZUGUCHI, K ;
SHIMONO, T ;
DAMM, G ;
GLUECK, J ;
TAKATANI, S ;
NOON, GP ;
NOSE, Y .
ARTIFICIAL ORGANS, 1994, 18 (01) :17-24
[8]  
Parnis SM, 1997, ASAIO J, V43, pM576
[9]  
Sakuma I, 1993, ASAIO J, V39, pM433
[10]  
Takami Y, 1997, ARTIF ORGANS, V21, P1132