Determination of tungsten in niobium-tantalum, vanadium and molybdenum bearing geological samples using derivative spectrophotometry and ICP-AES

被引:24
作者
Padmasubashini, V [1 ]
Ganguly, MK [1 ]
Satyanarayana, K [1 ]
Malhotra, RK [1 ]
机构
[1] Dept Atom Energy, Chem Lab, Atom Directorate Explorat & Res, Hyderabad 500016, Andhra Pradesh, India
关键词
tungsten; niobate-tantalates; geological materials; derivative spectrophotometry; inductively coupled plasma atomic emission spectrometry;
D O I
10.1016/S0039-9140(99)00165-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Two different procedures, one using derivative spectrophotometry and another using inductively coupled plasma atomic emission spectrometry (ICP-AES) have been developed for the determination of tungsten in niobate-tantalates, tin slag samples, ores, concentrates and vanadium and molybdenum bearing geological materials. In the first method involving derivative spectrophotometry, 0.05-0.5 g of the sample is fused with sodium hydroxide, the tungsten is extracted by leaching the melt with distilled water and estimated as thiocyanate using a second derivative spectrophotometric method in the presence of interferents, i.e. Nb, Mo and V, without separating them. Mixtures of tungsten with V, Nb and Mo are used for standardizing the various parameters like zero-crossing wavelength, wavelength range, etc. Tolerance limits for V, Nb and Mo have also been evaluated. In the second method involving ICP-AES, 0.05-0.5 g of sample is fused with KHSO4 to a clear melt and dissolved in ammonium oxalate solution. Ammonium hydroxide precipitation is then carried out to separate Nb and Ta as hydroxides and the filtrate is boiled with nitric acid to destroy the oxalates before aspiration into the plasma for measurement of tungsten values by ICP-AES using the 207.911 nm emission line. Both methods have been applied to niobate-tantalate and tin slag samples and the results obtained are reported in this paper. The values obtained by both methods are in good agreement with each other. The proposed methods have also been applied to the determination of tungsten in two Canadian Certified Reference Standards (CT-1 and MP-2) and the values obtained are in good agreement with the certified values and the R.S.D.% in case of the ICP-AES method varied from 1-2% at > 1000 mu g g(-1) level to 9.4% at the 20 mu g g(-1) level whereas the R.S.D.% in case of the derivative method varied from 1 to 7.8%. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:669 / 676
页数:8
相关论文
共 21 条
[1]  
BHATTACHARJEE S, 1993, SPECTROSC EUR, V4, P26
[2]  
BOWMAN WS, 1990, CCRMP901E, P63
[3]   THE SPECTROCHEMICAL (ICP-AES) DETERMINATION OF TUNGSTEN IN TUNGSTEN ORES, CONCENTRATES, AND ALLOYS - AN EVALUATION AS AN ALTERNATIVE TO THE CLASSICAL GRAVIMETRIC PROCEDURE [J].
BRENNER, IB ;
ERLICH, S .
APPLIED SPECTROSCOPY, 1984, 38 (06) :887-890
[4]   Estimation of tungsten in complex matrices using toluene 3,4-dithiol and thiocyanate [J].
Chattopadhyay, P ;
Mistry, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1998, 360 (01) :38-42
[5]   SEQUENTIAL DETERMINATION OF MOLYBDENUM AND TUNGSTEN IN SILICATE ROCKS BY A SPECTROPHOTOMETRIC METHOD [J].
FAN, ZH ;
ZHANG, LX ;
ZHOU, HD .
ANALYTICA CHIMICA ACTA, 1992, 270 (01) :267-269
[6]   SPECTROPHOTOMETRIC DETERMINATION OF TUNGSTEN WITH THIOCYANATE .1. A REVIEW OF PROCEDURES [J].
FOGG, AG ;
MARRIOTT, DR ;
BURNS, DT .
ANALYST, 1970, 95 (1135) :848-&
[7]  
GANGULY MK, 1989, P NATL WORKSH TUNGST, P139
[8]   DETERMINATION OF W AND MO IN NATURAL SPRING WATERS BY ICP-AES (INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY) AND ICP-MS (INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY) - APPLICATION TO SOUTH NAHANNI RIVER AREA, NWT, CANADA [J].
HALL, GEM ;
JEFFERSON, CW ;
MICHEL, FA .
JOURNAL OF GEOCHEMICAL EXPLORATION, 1988, 30 (01) :63-84
[9]   DETERMINATION OF LOW CONCENTRATIONS OF TUNGSTEN AND MOLYBDENUM IN GEOLOGICAL-MATERIALS USING INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY WITH PRECONCENTRATION ON ACTIVATED-CHARCOAL [J].
HALL, GEM ;
PELCHAT, JC ;
DESILVA, KN .
ANALYST, 1987, 112 (05) :631-635
[10]   DETERMINATION OF TUNGSTEN AND MOLYBDENUM AT LOW-LEVELS IN GEOLOGICAL-MATERIALS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
HALL, GEM ;
PARK, CJ ;
PELCHAT, JC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (02) :189-196