摘要
This work for the first time presents the full integration and automation concept for a range of bioassays leveraged by cascading a centrifugo-pneumatic valving scheme to sequentially move several liquids through shared channel segments for multi-step sample preparation into the detection zone. This novel centrifugo-pneumatic liquid handling significantly simplifies system manufacture by obviating the need for complex surface functionalization procedures or hybrid material integration, as it is common in conventional valving methods such as capillary burst valves or sacrificial valves. Based on the centrifugo-pneumatic valving scheme, this work presents a toolkit of operational elements implementing liquid loading/transfer, metering, mixing and sedimentation in a microstructured polymer disc. As a proof of concept for the broad class of homogeneous bioassays, the full integration and automation of a colorimetric nitrate/nitrite test for the detection of clinically relevant nitric oxide (NO) in whole blood is implemented. First, 40 mu L of plasma is extracted from a 100 mu L sample of human blood, incubated for one hour with the enzymatic mixture (60 mu L), and finally reacted with 100 mu L of colorimetric (Greiss) reagents. Following just a single loading phase at the beginning of the process, all of these steps are automated through the centrifugo-pneumatic cascade with a high level of flow control and synchronization. Our system shows good correlation with controls up to 50 mu M of nitrate, which adequately covers the healthy human range (4 to 45.3 mu M).
引用
收藏
页码:685 / 694
页数:10
相关论文
共 35 条
[1]
Thermo-pneumatic pumping in centrifugal microfluidic platforms [J].
Abi-Samra, Kameel ;
Clime, Liviu ;
Kong, Ling ;
Gorkin, Robert, III ;
Kim, Tae-Hyeong ;
Cho, Yoon-Kyoung ;
Madou, Marc .
MICROFLUIDICS AND NANOFLUIDICS, 2011, 11 (05) :643-652
[2]
Polymer microfabrication technologies for microfluidic systems [J].
Becker, Holger ;
Gaertner, Claudia .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2008, 390 (01) :89-111
[3]
Optical sensing system based on wireless paired emitter detector diode device and ionogels for lab-on-a-disc water quality analysis [J].
Czugala, Monika ;
Gorkin, Robert, III ;
Phelan, Thomas ;
Gaughran, Jennifer ;
Curto, Vincenzo Fabio ;
Ducree, Jens ;
Diamond, Dermot ;
Benito-Lopez, Fernando .
LAB ON A CHIP, 2012, 12 (23) :5069-5078
[4]
The centrifugal microfluidic bio-disk platform [J].
Ducree, Jens ;
Haeberle, Stefan ;
Lutz, Sascha ;
Pausch, Sarah ;
von Stetten, Felix ;
Zengerle, Roland .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2007, 17 (07) :S103-S115
[5]
Liquid recirculation in microfluidic channels by the interplay of capillary and centrifugal forces [J].
Garcia-Cordero, Jose L. ;
Basabe-Desmonts, Lourdes ;
Ducree, Jens ;
Ricco, Antonio J. .
MICROFLUIDICS AND NANOFLUIDICS, 2010, 9 (4-5) :695-703
[6]
Endothelial Dysfunction in Patients with Noncomplicated and Complicated Hypertension [J].
Goch, Aleksander ;
Banach, Maciej ;
Mikhailidis, Dimitri P. ;
Rysz, Jacek ;
Henryk Goch, Jan .
CLINICAL AND EXPERIMENTAL HYPERTENSION, 2009, 31 (01) :20-30
[7]
Gorkin R., 2011, 16 INT C SOL STAT SE
[8]
Centrifugal microfluidics for biomedical applications [J].
Gorkin, Robert ;
Park, Jiwoon ;
Siegrist, Jonathan ;
Amasia, Mary ;
Lee, Beom Seok ;
Park, Jong-Myeon ;
Kim, Jintae ;
Kim, Hanshin ;
Madou, Marc ;
Cho, Yoon-Kyoung .
LAB ON A CHIP, 2010, 10 (14) :1758-1773
[9]
Pneumatic pumping in centrifugal microfluidic platforms [J].
Gorkin, Robert, III ;
Clime, Liviu ;
Madou, Marc ;
Kido, Horacio .
MICROFLUIDICS AND NANOFLUIDICS, 2010, 9 (2-3) :541-549
[10]
Sensitivity enhancement for colorimetric glucose assays on whole blood by on-chip beam-guidance [J].
Grumann, M. ;
Steigert, J. ;
Riegger, L. ;
Moser, I. ;
Enderle, B. ;
Riebeseel, K. ;
Urban, G. ;
Zengerle, R. ;
Ducree, J. .
BIOMEDICAL MICRODEVICES, 2006, 8 (03) :209-214