Brain tissue pO(2)-monitoring: Catheter stability and complications

被引:46
作者
Dings, J
Meixensberger, J
Roosen, K
机构
[1] Neurosurgical Department, University of Würzburg, Würzburg
[2] Department of Neurosurgery, University of Würzburg, 97080 Würzburg
关键词
brain tissue pO(2); complications; brain oxygenation; head injury; subarachnoid hemorrhage; catheter stability;
D O I
10.1080/01616412.1997.11740806
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The authors report on the stability and complications of 73 LICOX brain ti-pO(2)-microcatheters in 70 patients. Mean monitoring time was 7.5 +/- 4.0 days. Patients prone to cerebral hypoxia (after severe head injury (GCS < 9) or a subarachnoid hemorrhage) had a ti-pO(2)-microcatheter inserted next to the ICP-probe in the typical frontal position. After the first 15 insertions, instead of the 3-way-screw (needing a 6 mm burrhole), a I-way-screw (needing a 2.7 mm burrhole) was used for fixation in the bone; by doing so, the procedure can be performed in the ICU and takes only 15 min. Whenever possible a calibration at room air (to determine the sensitivity-drift) and in oxygen free solution (to determine the zero-drift) was performed after removal of the catheters. Ideally the expected pO(2) at room air was around 154 mmHg (temperature dependent) and at zero calibration 0 mmHg. Mean sensitivity-drift for 54 catheters was -8.5 +/- 15.4%. Dividing the catheters into groups, depending on the duration of monitoring (1-4, 5-8 and 9-16 days), revealed that the greatest part of the (negative) sensitivity-drift occurred during day 1-4 after insertion. After 1 week of monitoring sometimes a positive drift occurred (being far less than the negative drift during the first 4 days). Compared to the old catheters (-10.3 +/- 17.3%) ion the first half of the patients) the new ones showed a lower sensitivity-drift (-6.8 +/- 13.4%). The zero-drift of 56 catheters was low with mean drift after 7.5 +/- 4.0 days of 1.5 +/- 1.5 mmHg. Here also the highest drift occurred on day 1-4 after insertion. No infection was seen and 2 times (2.7%) a small hematoma, not needing evacuation occurred. As the ti-pO(2)-catheter (having a smaller diameter) and the ICP-catheter were inserted at the same time, one cannot distinguish which catheter caused the hematoma. A possible explanation for the occurrence of the two hematomas is the insertion of the catheters too close to the midline. The authors conclude that LICOX ti-pO(2)-monitoring is a safe and reliable method. Further decrease of the complication rate and increase of the catheter-stability may be expected.
引用
收藏
页码:241 / 245
页数:5
相关论文
共 13 条
  • [1] ULTRA-EARLY EVALUATION OF REGIONAL CEREBRAL BLOOD-FLOW IN SEVERELY HEAD-INJURED PATIENTS USING XENON-ENHANCED COMPUTERIZED-TOMOGRAPHY
    BOUMA, GJ
    MUIZELAAR, JP
    STRINGER, WA
    CHOI, SC
    FATOUROS, P
    YOUNG, HF
    [J]. JOURNAL OF NEUROSURGERY, 1992, 77 (03) : 360 - 368
  • [2] CEREBRAL-CIRCULATION AND METABOLISM AFTER SEVERE TRAUMATIC BRAIN INJURY - THE ELUSIVE ROLE OF ISCHEMIA
    BOUMA, GJ
    MUIZELAAR, JP
    CHOI, SC
    NEWLON, PG
    YOUNG, HF
    [J]. JOURNAL OF NEUROSURGERY, 1991, 75 (05) : 685 - 693
  • [3] CONTINUOUS POSTOPERATIVE MONITORING OF CORTICAL BLOOD-FLOW AND INTRACRANIAL-PRESSURE
    CARTER, LP
    GRAHM, T
    BAILES, JE
    BICHARD, W
    SPETZLER, RF
    [J]. SURGICAL NEUROLOGY, 1991, 35 (01): : 36 - 39
  • [4] DEARDEN NM, 1993, ACTA NEUROCHIR, P91
  • [5] Luerssen T.G.S.P., 1989, INTRACRANIAL PRESSUR, P35
  • [6] ACUTE REGIONAL CEREBRAL BLOOD-FLOW CHANGES CAUSED BY SEVERE HEAD-INJURIES
    MARION, DW
    DARBY, J
    YONAS, H
    [J]. JOURNAL OF NEUROSURGERY, 1991, 74 (03) : 407 - 414
  • [7] METZ C, 1993, ANASTH INTENSIVMED, V34, P345
  • [8] INTRA-CRANICAL PRESSURE - TO MONITOR OR NOT TO MONITOR - A REVIEW OF OUR EXPERIENCE WITH SEVERE HEAD-INJURY
    NARAYAN, RK
    KISHORE, PRS
    BECKER, DP
    WARD, JD
    ENAS, GG
    GREENBERG, RP
    DASILVA, AD
    LIPPER, MH
    CHOI, SC
    MAYHALL, CG
    LUTZ, HA
    YOUNG, HF
    [J]. JOURNAL OF NEUROSURGERY, 1982, 56 (05) : 650 - 659
  • [9] Piek J, 1994, Zentralbl Neurochir, V55, P48