摘要
Background It is known that first-generation quantitative coronary angiography (QCA) systems overestimate small vessel sizes owing to the point-spread function of the respective X-ray imaging chain, With second-generation systems new algorithms were introduced to correct for this source of error. Objective To evaluate the efficiency of the modified contour detection algorithms. Methods Six second-generation QCA systems (CMS, QANSAD, AWOS, CAAS II, Cardio 500, and Angioimage) were validated and compared with first-generation systems (CAAS and ARTREK). By using an arterial phantom consisting of stenotic and nonstenotic glass tubes (of diameters 0.5-5.0 mm) the accuracy and precision of each analysis system, as well as their additional accuracy and precision values for phantom diameters less than or equal to 1.0 mm were determined. Results All systems had high accuracy and precision values, but first-generation systems overestimated small vessel diameters. With second-generation systems a significantly improved accuracy in the submillimeter range (an accuracy within +/- 0.028 mm) was obtained, This improvement was accompanied by a moderate reduction in precision in the submillimeter range. Conclusion The new algorithms of the second-generation QCA systems allow accurate and reliable measurements of small coronary dimensions and, therefore, precise analysis of coronary stenoses of moderate-to-high grade seems feasible with the improved accuracy of the new systems.
引用
收藏
页码:83 / 90
页数:8
相关论文
共 40 条
[1]   CHANGE IN DIAMETER OF CORONARY-ARTERY SEGMENTS ADJACENT TO STENOSIS AFTER PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - FAILURE OF PERCENT DIAMETER STENOSIS MEASUREMENT TO REFLECT MORPHOLOGIC CHANGES INDUCED BY BALLOON DILATION [J].
BEATT, KJ ;
LUIJTEN, HE ;
DEFEYTER, PJ ;
VANDENBRAND, M ;
REIBER, JHC ;
SERRUYS, PW ;
TENKATEN, HJ ;
VANES, GA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 12 (02) :315-323
[2]   RESTENOSIS AFTER CORONARY ANGIOPLASTY - THE PARADOX OF INCREASED LUMEN DIAMETER AND RESTENOSIS [J].
BEATT, KJ ;
SERRUYS, PW ;
LUIJTEN, HE ;
RENSING, BJ ;
SURYAPRANATA, H ;
DEFEYTER, P ;
VANDENBRAND, M ;
LAARMAN, GJ ;
ROELANDT, J ;
VANES, GA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (02) :258-266
[3]   RESTENOSIS AFTER CORONARY ANGIOPLASTY - NEW STANDARDS FOR CLINICAL-STUDIES [J].
BEATT, KJ ;
SERRUYS, PW ;
HUGENHOLTZ, PG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (02) :491-498
[4]  
BEAUMAN GJ, 1994, PROGR QUANTITATIVE C, P87
[5]  
BEIER J, 1993, AUTOMATISCHE QUANTIT, P1
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]   QUANTITATIVE CORONARY ARTERIOGRAPHY - ESTIMATION OF DIMENSIONS, HEMODYNAMIC RESISTANCE, AND ATHEROMA MASS OF CORONARY-ARTERY LESIONS USING ARTERIOGRAM AND DIGITAL COMPUTATION [J].
BROWN, BG ;
BOLSON, E ;
FRIMER, M ;
DODGE, HT .
CIRCULATION, 1977, 55 (02) :329-337
[8]   QUANTITATIVE CORONARY ANGIOGRAPHY TO MEASURE PROGRESSION AND REGRESSION OF CORONARY ATHEROSCLEROSIS - VALUE, LIMITATIONS, AND IMPLICATIONS FOR CLINICAL-TRIALS [J].
DEFEYTER, PJ ;
SERRUYS, PW ;
DAVIES, MJ ;
RICHARDSON, P ;
LUBSEN, J ;
OLIVER, MF .
CIRCULATION, 1991, 84 (01) :412-423
[9]   A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE [J].
FISCHMAN, DL ;
LEON, MB ;
BAIM, DS ;
SCHATZ, RA ;
SAVAGE, MP ;
PENN, I ;
DETRE, K ;
VELTRI, L ;
RICCI, D ;
NOBUYOSHI, M ;
CLEMAN, M ;
HEUSER, R ;
ALMOND, D ;
TEIRSTEIN, PS ;
FISH, RD ;
COLOMBO, A ;
BRINKER, J ;
MOSES, J ;
SHAKNOVICH, A ;
HIRSHFELD, J ;
BAILEY, S ;
ELLIS, S ;
RAKE, R ;
GOLDBERG, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) :496-501
[10]   PITFALLS IN THE DETERMINATION OF ABSOLUTE DIMENSIONS USING ANGIOGRAPHIC CATHETERS AS CALIBRATION DEVICES IN QUANTITATIVE ANGIOGRAPHY [J].
FORTIN, DF ;
SPERO, LA ;
CUSMA, JT ;
SANTORO, L ;
BURGESS, R ;
BASHORE, TM .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 68 (11) :1176-1182