Analysis of artefacts and detail resolution of lung MRI with breath-hold T1-weighted gradient-echo and T2-weighted fast spin-echo sequences with respiratory triggering

被引:40
作者
Biederer, J
Reuter, M
Both, M
Muhle, C
Grimm, J
Graessner, J
Heller, M
机构
[1] Univ Hosp Kiel, Dept Diagnost Radiol, D-24105 Kiel, Germany
[2] Univ Hosp Kiel, Dept Nucl Med, D-24105 Kiel, Germany
[3] Siemens Aktiengesell, Hamburg, Germany
关键词
MRI; lung; image-quality rating;
D O I
10.1007/s00330-001-1147-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to evaluate feasibility and limitations of two MR sequences for imaging of the lung using a semi-quantitative rating scale. Ten healthy volunteers were assessed with a breath-hold T1-weighted gradient-recalled-echo (TR/TE=129/2.2 ms. matrix 173X256) and a T2-weighted turbo spin-echo (TSE) sequence with respiratory triggering (TR/TE=3000-5000/126 ms. matrix 270x512) in axial 6-mm slices. The T1-weighted GRE protocol included a pre-saturation Pulse over the mediastinal structures. Artefacts and resolution of vessel/airway structures in each lung segment were evaluated by two observers ( 10 volunteers, 186 segments). Cardiac and vessel pulsation artefacts predominated on T1-weighted GRE, respiration artefacts on T2-weighted TSE (lingula and middle lobe). Pre-saturation of the mediastinum reduced pulsation artefacts on T1-weighted GRE. T1-weighted GRE images were improved by bright flow signal of vessels, whereas image quality of T2-weighted TSE was reduced by black-blood effects in central parts of the lung. Delineation of lung periphery and the mediastinum was superior with T2-weighted TSE. Segmental/sub-segmental vessels (up to fourth/fifth order) and bronchi (up to third order) were identified. All 180 lung, segments were imaged in diagnostic quality with at least one of the two sequences (T1-weighted GRE not diagnostic in 9 of 180, T2-weighted TSE in 4 of 180). Both sequences were found to be complementary: superior identification of gross lung anatomy with T1-weighted GRE and higher detail resolution in the periphery and the mediastinum with T2-weighted TSE.
引用
收藏
页码:378 / 384
页数:7
相关论文
共 31 条
[1]   MULTISLICE, BREATHHOLD IMAGING OF THE LUNG WITH SUBMILLISECOND ECHO TIMES [J].
ALSOP, DC ;
HATABU, H ;
BONNET, M ;
LISTERUD, J ;
GEFTER, W .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) :678-682
[2]  
Amundsen T, 2000, J Magn Reson Imaging, V12, P224, DOI 10.1002/1522-2586(200008)12:2<224::AID-JMRI3>3.0.CO
[3]  
2-E
[4]  
BIEDERER J, 2000, EUR RADIOL, V10, pG7
[5]  
BIEDERER J, 2000, EUR RADIOL, V10, pG8
[6]  
BIEDERER J, 2001, EUR RADIOL, V11, P244
[7]  
BITTNER RC, 1999, IMAGING DECISIONS, V3, P2
[8]   Fast magnetic resonance imaging techniques [J].
Chen, Q ;
Stock, KW ;
Prasad, PV ;
Hatabu, H .
EUROPEAN JOURNAL OF RADIOLOGY, 1999, 29 (02) :90-100
[9]  
Chen Q, 1998, MAGN RESON MATER PHY, V7, P153
[10]   Detection of emphysema in rat lungs by using magnetic resonance measurements of 3He diffusion [J].
Chen, XJ ;
Hedlund, LW ;
Möller, HE ;
Chawla, MS ;
Maronpot, RR ;
Johnson, GA .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2000, 97 (21) :11478-11481