Performance of a Dual Sample Introduction System with Conventional Concentric Nebulizers for Simultaneous Determination of Hydride and Non-Hydride Forming Elements by ICP-OES

被引:16
作者
Benzo, Zully [1 ]
Maldonado, Domingo [1 ]
Chirinos, Jose [2 ]
Marcano, Eunice [1 ]
Gomez, Clara [1 ]
Quintal, Manuelita [1 ]
Salas, Janeth [1 ]
机构
[1] Inst Venezolano Invest Cient, Ctr Quim, Lab Quim Analit, Caracas 1020A, Venezuela
[2] Cent Univ Venezuela, Escuela Quim, Ctr Quim Analit, Caracas, Venezuela
关键词
Dual-mode introduction system; Hydride generation; ICP-OES;
D O I
10.1080/10739140802448275
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A dual sample introduction system that combines the benefits of nebulization and vapour generation in a single device is described. It consists of two commercial conventional concentric nebulizers coupled to a modified cyclonic chamber. The effect(s) of the solvent load produced by the amount of liquid carried for the system by the two nebulizer assemblies is investigated. Better sensitivities, similar precision and DL's (with the exception of hydride forming elements) were obtained compared with those obtained with the system working in single mode. Long term stability was less than 7% with the dual mode, being 2% and 6% for the non-hydride and hydride forming elements, respectively, in the single mode operation. DL's obtained are of the same order of magnitude as those reported by several authors, with the exception of Se whose conditions were in compromise with the optimal reached for the rest of the elements. Accuracy of the dual system was proved by analyzing NIST 1648, urban particulate matter, with satisfactory results.
引用
收藏
页码:598 / 610
页数:13
相关论文
共 23 条
[1]   Simultaneous determination of hydride (Se) and non-hydride-forming (Ca, Mg, K, P, S and Zn) elements in various beverages (beer, coffee, and milk), with minimum sample preparation, by ICP-AES and use of a dual-mode sample-introduction system [J].
Asfaw, A ;
Wibetoe, G .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2005, 382 (01) :173-179
[2]   Dual mode sample introduction for multi-element determination by ICP-MS:: the optimization and use of a method based on simultaneous introduction of vapor formed by NaBH4 reaction and aerosol from the nebulizer [J].
Asfaw, Alemayehu ;
Wibetoe, Grethe .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2006, 21 (10) :1027-1035
[3]   DETECTION LIMITS AND SPECTRAL INTERFERENCES IN ATOMIC-EMISSION SPECTROMETRY [J].
BOUMANS, PWJM .
ANALYTICAL CHEMISTRY, 1994, 66 (08) :A459-A467
[4]   Minimization of transition metal interferences with hydride generation techniques [J].
Ding, WW ;
Sturgeon, RE .
ANALYTICAL CHEMISTRY, 1997, 69 (03) :527-531
[5]   Dual nebulizer sample introduction system for simultaneous determination of volatile elemental hydrides and other elements [J].
Gomez, Luis R. ;
Marquez, Gracybell D. ;
Chirinos, Jose R. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 386 (01) :188-195
[6]   A NEW NEBULIZER-HYDRIDE GENERATOR SYSTEM FOR SIMULTANEOUS MULTIELEMENT INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
HUANG, BL ;
ZHANG, ZY ;
ZENG, XJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (1-2) :129-137
[7]   RECENT STUDIES ON INDUCTIVELY COUPLED PLASMA SAMPLE INTRODUCTION .1. HYDRIDE GENERATION [J].
HUANG, BL ;
ZENG, XJ ;
ZHANG, ZY ;
LIU, JA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1988, 43 (4-5) :381-389
[8]   A SIMPLE VERSATILE HYDRIDE-GENERATION CONFIGURATION FOR INDUCTIVELY COUPLED PLASMAS [J].
HUTTON, RC ;
PRESTON, B .
ANALYST, 1983, 108 (1292) :1409-1411
[9]   HYDRIDE GENERATOR SYSTEM FOR A 1-KW INDUCTIVELY COUPLED PLASMA [J].
HWANG, JD ;
GUENTHER, GD ;
DIOMIGUARDI, JP .
ANALYTICAL CHEMISTRY, 1989, 61 (03) :285-288
[10]   Analytical evaluation of a dual micronebulizer sample introduction system for inductively coupled plasma spectrometry [J].
Maldonado, Domingo ;
Chirinos, Jose ;
Benzo, Zully ;
Gomez, Clara ;
Marcano, Eunice .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2006, 21 (08) :743-749