Implementation and evaluation of a 3D one-step late reconstruction algorithm for 3D positron emission tomography brain studies using median root prior

被引:31
作者
Bettinardi, V
Pagani, E
Gilardi, MC
Alenius, S
Thielemans, K
Teras, M
Fazio, F
机构
[1] Univ Milano Bicocca, INB CNR, Sci Inst HS Raffaele, Dept Nucl Med, I-20132 Milan, Italy
[2] Tampere Univ Technol, Signal Proc Lab, FIN-33101 Tampere, Finland
[3] Turku PET Ctr, Turku, Finland
[4] Imaging Res Solut Ltd, Hammersmith Hosp, London, England
关键词
iterative reconstruction algorithm; 3D; positron emission tomography;
D O I
10.1007/s002590100651
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A fully three-dimensional (3D) one-step late (OSL), maximum a posteriori (MAP) reconstruction algorithm based on the median root prior (MRP) was implemented and evaluated for the reconstruction of 3D positron emission tomography (PET) studies. The algorithm uses the ordered subsets (OS) scheme for convergence acceleration and data update during iterations. The algorithm was implemented using the software package developed within the EU project PARAPET (www.brunel.ac.uk/similar tomasrppet). The MRP algorithm was evaluated using experimental phantom and real 3D PET brain studies. Various experimental set-ups in terms of activity distribution and counting statistics were considered. The performance of the algorithm was assessed by calculating figures of merit such as: contrast, coefficient of variation, activity ratio between two regions and full width at half of maximum for resolution measurements. The performance of MRP was compared with that of 3D ordered subsets-expectation maximisation (OSEM) and 3D re-projection (3DRP) algorithms. In all the experimental situations considered, MRP showed: (1) convergence to a stable solution, (2) effectiveness in noise reduction, particularly for low statistics data, (3) good preservation of spatial details. Compared with the OSEM and 3DRP algorithms, MRP provides comparable or better results depending on the parameters used for the reconstruction of the images.
引用
收藏
页码:7 / 18
页数:12
相关论文
共 24 条
[1]   Attenuation correction for PET using count-limited transmission images reconstructed with median root prior [J].
Alenius, S ;
Ruotsalainen, U ;
Astola, J .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (03) :646-651
[2]  
ALENIUS S, TAMPERE U TECHNOLOGY, V265
[3]  
ALENIUS S, 2000, IEEE NSS MIC C
[4]   An automatic classification technique for attenuation correction in positron emission tomography [J].
Bettinardi, V ;
Pagani, E ;
Gilardi, MC ;
Landoni, C ;
Riddell, C ;
Rizzo, G ;
Castiglioni, I ;
Belluzzo, D ;
Lucignani, G ;
Schubert, S ;
Fazio, F .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (05) :447-458
[5]   ITERATIVE METHODS FOR IMAGE DEBLURRING [J].
BIEMOND, J ;
LAGENDIJK, RL ;
MERSEREAU, RM .
PROCEEDINGS OF THE IEEE, 1990, 78 (05) :856-883
[6]   A unified approach to statistical tomography using coordinate descent optimization [J].
Bouman, CA ;
Sauer, K .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 1996, 5 (03) :480-492
[7]   PRECISION AND ACCURACY OF REGIONAL RADIOACTIVITY QUANTITATION USING THE MAXIMUM-LIKELIHOOD EM RECONSTRUCTION ALGORITHM [J].
CARSON, RE ;
YAN, YC ;
CHODKOWSKI, B ;
YAP, TK ;
DAUBEWITHERSPOON, ME .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (03) :526-537
[8]   BAYESIAN RECONSTRUCTIONS FROM EMISSION TOMOGRAPHY DATA USING A MODIFIED EM ALGORITHM [J].
GREEN, PJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (01) :84-93
[9]   ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA [J].
HUDSON, HM ;
LARKIN, RS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :601-609
[10]   Enhanced 3D PET OSEM reconstruction using inter-update Metz filtering [J].
Jacobson, M ;
Levkovitz, R ;
Ben-Tal, A ;
Thielemans, K ;
Spinks, T ;
Belluzzo, D ;
Pagani, E ;
Bettinardi, V ;
Gilardi, MC ;
Zverovich, A ;
Mitra, G .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (08) :2417-2439