Statistics of the log-compressed echo envelope

被引:46
作者
Dutt, V
Greenleaf, JF
机构
[1] Biodynamics Research Unit, Mayo Clinic and Foundation, Rochester
关键词
D O I
10.1121/1.414999
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Log compression of A lines to produce B-scan images in clinical ultrasound imaging systems is a standard procedure to control the dynamic range of the images. The statistics of such compressed images in terms of underlying scatterer statistics have not been derived. The statistics are analyzed for partially formed speckle using a general K distribution model of envelope statistics to derive the density function for the log-compressed envelope. This density function is used to elucidate the relation between the moments of the compressed envelope, the compression parameters, and the statistics of the scatterers. The analysis shows that the mean of the log-compressed envelope is an increasing function of both the backscattered energy and the effective scatterer density. The variance of the log-compressed envelope is a decreasing function of the effective scatterer density and is independent of the backscattered energy. (C) 1996 Acoustical Society of America.
引用
收藏
页码:3817 / 3825
页数:9
相关论文
共 15 条
[1]   Speckle analysis using signal to noise ratios based on fractional order moments [J].
Dutt, V ;
Greenleaf, JF .
ULTRASONIC IMAGING, 1995, 17 (04) :251-268
[2]  
Gradshteyn I. S., 1994, TABLES INTEGRALS SER
[3]  
HOCHSTADT H, 1986, FUNCTIONS MATH PHYSI
[4]   GENERALIZED K-DISTRIBUTION - A STATISTICAL-MODEL FOR WEAK SCATTERING [J].
JAKEMAN, E ;
TOUGH, RJA .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (09) :1764-1772
[5]   MODEL FOR NON-RAYLEIGH SEA ECHO [J].
JAKEMAN, E ;
PUSEY, PN .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 1976, 24 (06) :806-814
[6]   STATISTICS OF K-DISTRIBUTED NOISE [J].
JAKEMAN, E .
JOURNAL OF PHYSICS A-MATHEMATICAL AND GENERAL, 1980, 13 (01) :31-48
[7]  
KAPLAN D, 1993, ULTRASON, P961, DOI 10.1109/ULTSYM.1993.339648
[8]  
MOLTHEN RC, 1993, ULTRASON, P957, DOI 10.1109/ULTSYM.1993.339649
[9]   NON-RAYLEIGH STATISTICS OF ULTRASONIC BACKSCATTERED SIGNALS [J].
NARAYANAN, VM ;
SHANKAR, PM ;
REID, JM .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1994, 41 (06) :845-852
[10]   USE OF NON-RAYLEIGH STATISTICS FOR THE IDENTIFICATION OF TUMORS IN ULTRASONIC B-SCANS OF THE BREAST [J].
SHANKAR, PM ;
REID, JM ;
ORTEGA, H ;
PICCOLI, CW ;
GOLDBERG, BB .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1993, 12 (04) :687-692