Organ perfusion with hemopump device assistance with and without intraaortic balloon pumping

被引:20
作者
Meyns, B [1 ]
Nishimura, Y [1 ]
Racz, R [1 ]
Jashari, R [1 ]
Flameng, W [1 ]
机构
[1] KATHOLIEKE UNIV LEUVEN,UNIV HOSP GASTHUISBERG,DEPT CARDIAC SURG,B-3001 LOUVAIN,BELGIUM
关键词
D O I
10.1016/S0022-5223(97)70152-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Our objective was to analyze the potential advantage of combining an intraaortic balloon pump with a transthoracic Hemopump device (Medtronic, Inc., Minneapolis, Minn.) (Nimbus Medical, Inc., Rancho Cordova, Calif.). Methods: Twelve sheep underwent implantation of a transthoracic Hemopump device and an intraaortic balloon pump. In the first series (n = 6), we analyzed the influence of the counterpulsation on the performance of the Hemopump device. In the second group (n = 6), hemodynamic changes, myocardial wall thickening, organ perfusion, and myocardial perfusion (determined with colored microspheres) were analyzed under the following conditions: (1) control situation, (2) during application of coronary stenosis, (3) during support with the Hemopump device, and (4) during support with the Hemopump device combined with intraaortic balloon pump support. Results: In the first series, we found that addition of counterpulsation reduced output with the Hemopump device by 11.1% +/- 6%. In the second series, it was shown that coronary stenosis significantly reduced contractility (rate of pressure change and wall thickening) but did not cause hemodynamic collapse. Myocardial blood Bow was significantly reduced in the poststenotic subendocardial regions (mean subendocardial blood flow dropped from 78 +/- 33 to 24 +/- 17 ml/min/100 gm; p = 0.0486). Support with the Hemopump device alone improved the ratio of subendocardial to subepicardial blood flow, but endocardial underperfusion remained (analysis of variance, p < 0.001). The Hemopump device with an intraaortic balloon pump completely restored perfusion in poststenotic regions. Peripheral organ perfusion did not change during ischemia or mechanical support. Conclusions: The association of balloon counterpulsation with the Hemopump device reduces the Hemopump output by 11% and increases myocardial blood flow to ischemic regions. Perfusion to peripheral organs remains unaltered. The transthoracic Hemopump device combined with an intraaortic balloon pump is an ideal support system for the ischemic, failing heart.
引用
收藏
页码:243 / 253
页数:11
相关论文
共 13 条
[1]   VASCULAR COMPLICATIONS FROM INTRAAORTIC BALLOONS - RISK ANALYSIS [J].
BARNETT, MG ;
SWARTZ, MT ;
PETERSON, GJ ;
NAUNHEIM, KS ;
PENNINGTON, DG ;
VACA, KJ ;
FIORE, AC ;
MCBRIDE, LR ;
PEIGH, P ;
WILLMAN, VL ;
KAISER, GC .
JOURNAL OF VASCULAR SURGERY, 1994, 19 (01) :81-89
[2]   MEASUREMENT OF REGIONAL MYOCARDIAL BLOOD-FLOW WITH MULTIPLE COLORED MICROSPHERES [J].
KOWALLIK, P ;
SCHULZ, R ;
GUTH, BD ;
SCHADE, A ;
PAFFHAUSEN, W ;
GROSS, R ;
HEUSCH, G .
CIRCULATION, 1991, 83 (03) :974-982
[3]  
MERHIGE ME, 1989, CIRCULATION, V80, P158
[4]   LEFT-VENTRICULAR ASSISTANCE WITH THE TRANSTHORACIC 24F HEMOPUMP FOR RECOVERY OF THE FAILING HEART [J].
MEYNS, BP ;
SERGEANT, PT ;
DAENEN, WJ ;
FLAMENG, WJ .
ANNALS OF THORACIC SURGERY, 1995, 60 (02) :392-397
[5]   DIASTOLIC BALLOON PUMPING (WITH CARBON DIOXIDE) IN AORTA-A MECHANICAL ASSISTANCE TO FAILING CIRCULATION [J].
MOULOPOULOS, SD ;
TOPAZ, S ;
KOLFF, WJ .
AMERICAN HEART JOURNAL, 1962, 63 (05) :669-&
[6]   COUNTERPULSATION - HISTORICAL BACKGROUND, TECHNICAL IMPROVEMENTS, HEMODYNAMIC AND METABOLIC EFFECTS [J].
NANAS, JN ;
MOULOPOULOS, SD .
CARDIOLOGY, 1994, 84 (03) :156-167
[7]   INTRAAORTIC BALLOON COUNTERPULSATION IN DENMARK 1988-1991 - EARLY RESULTS AND COMPLICATIONS [J].
OLSEN, PS ;
ARENDRUP, H ;
THIIS, JJ ;
KLAABORG, KE ;
HOLDGAARD, HO .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1993, 7 (12) :634-636
[8]   EFFECTS OF INTRA-AORTIC BALLOON COUNTERPULSATION ON CARDIAC PERFORMANCE, OXYGEN CONSUMPTION, AND CORONARY BLOOD FLOW IN DOGS [J].
POWELL, WJ ;
DAGGETT, WM ;
MAGRO, AE ;
BIANCO, JA ;
BUCKLEY, MJ ;
SANDERS, CA ;
KANTROWITZ, AR ;
AUSTEN, WG .
CIRCULATION RESEARCH, 1970, 26 (06) :753-+
[9]   CIRCULATION OF FETUS IN UTERO - METHODS FOR STUDYING DISTRIBUTION OF BLOOD FLOW CARDIAC OUTPUT AND ORGAN BLOOD FLOW [J].
RUDOLPH, AM ;
HEYMANN, MA .
CIRCULATION RESEARCH, 1967, 21 (02) :163-&
[10]   CHANGES IN HEMODYNAMICS AND CORONARY BLOOD-FLOW DURING LEFT-VENTRICULAR ASSISTANCE WITH THE HEMOPUMP [J].
SHIIYA, N ;
ZELINSKY, R ;
DELEUZE, PH ;
LOISANCE, DY .
ANNALS OF THORACIC SURGERY, 1992, 53 (06) :1074-1079