Radiation safety considerations for diagnostic radiology personnel

被引:55
作者
Brateman, L [1 ]
机构
[1] Univ Florida, Coll Med, Dept Radiol, Gainesville, FL 32610 USA
关键词
radiations; exposure to patients and personnel; protective and therapeutic agents and devices; radiology and radiologists; design of radiological facilities;
D O I
10.1148/radiographics.19.4.g99jl231037
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In radiation protection, the guiding philosophy is ALARA (as low as reasonably achievable), and states have regulatory authority. Dose limits are in part based on effective dose equivalent and differences in tissue sensitivities. In diagnostic radiology, the main source of occupational dose is scattered radiation from the patient-particularly from fluoroscopically guided procedures. Personnel stand near patients for long times, and angulated geometries with C-arm equipment may result in high personnel doses from backscatter. For all procedures, judicious applications of time, distance, and shielding affect dose. Appropriate use includes collimating properly, optimizing beam-on time, minimizing distances between image intensifier and patient, ensuring sufficient distance between patient and x-ray tube, and optimizing exposure rates for image quality and dose. Although dose limits typically regulate maximum whole-body dose, protective clothing worn by fluoroscopists reduces personnel risks; weighting factors can be applied to estimate effective dose equivalent. Pregnant personnel have lower limits, which apply only with voluntary declaration of pregnancy. With appropriate precautions, fetal doses can typically remain within recommended limits without changes in occupational tasks. Radiation workers in each state must ensure that regulations are appropriate. Then, for protection of both employees and employers, the rules can and must be followed.
引用
收藏
页码:1037 / 1055
页数:19
相关论文
共 26 条
[1]  
ACR, 1992, ACR B, V48, P21
[2]   THE IMPACT OF MATERNITY ON RADIOLOGISTS - THE AAWR POSITION AND ITS ACCEPTANCE BY WOMEN [J].
ADLER, YT ;
FERNBACH, SK ;
HAYMAN, LA ;
REDMAN, HC ;
RUMACK, C .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1986, 146 (02) :415-417
[3]  
American Association of Physicists in Medicine, 1998, 58 AAPM
[4]  
[Anonymous], ANN ICRP
[5]  
[Anonymous], ANN ICRP
[6]  
BALTER S, 1998, CATEGORICAL COURSE D, P223
[7]  
BUSHONG SC, 1984, RADIOLOGIC SCI TECHN, P545
[8]  
*C RAD CONTR PROGR, 1995, SUGG STAT REG CONTR, V1
[9]   ESTIMATION OF EFFECTIVE DOSE EQUIVALENT TO STAFF IN DIAGNOSTIC-RADIOLOGY [J].
FAULKNER, K ;
HARRISON, RM .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (01) :83-91
[10]  
*INT COMM RAD UN M, 1998, 60 ICRU INT COMM RAD