Relative dosimetry using active matrix flat-panel imager (AMFPI) technology

被引:94
作者
El-Mohri, Y [1 ]
Antonuk, LE [1 ]
Yorkston, J [1 ]
Jee, KW [1 ]
Maolinbay, M [1 ]
Lam, KL [1 ]
Siewerdsen, JH [1 ]
机构
[1] Univ Michigan, Med Ctr, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
关键词
relative dosimetry; active matrix flat-panel imager; electronic portal imager; amorphous silicon; quality assurance;
D O I
10.1118/1.598649
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The first examination of the use of active matrix flat-panel arrays for dosimetry in radiotherapy is reported. Such arrays are under widespread development for diagnostic and radiotherapy imaging. In the current study, an array consisting of 512X512 pixels with a pixel pitch of 508 mu m giving an area of 26X26 cm(2) has been used. Each pixel consists of a light sensitive amorphous silicon (a-Si:H) photodiode coupled to an a-Si:H thin-film transistor. Data was obtained from the array using a dedicated electronics system allowing real-time data acquisition. In order to examine the potential of such arrays as quality assurance devices for radiotherapy beams, field profile data at photon energies of 6 and 15 MV were obtained as a function of field size and thickness of overlying absorbing material (solid water). Two detection configurations using the array were considered: a configuration (similar to the imaging configuration) in which an overlying phosphor screen is used to convert incident radiation to visible light photons which are detected by the photodiodes; and a configuration without the screen where radiation is directly sensed by the photodiodes. Compared to relative dosimetry data obtained with an ion chamber, data taken using the former configuration exhibited significant differences whereas data obtained using the latter configuration was generally found to be in close agreement. Basic signal properties, which are pertinent to dosimetry, have been investigated through measurements of individual pixel response for fluoroscopic and radiographic array operation. For signal levels acquired within the first 25% of pixel charge capacity, the degree of linear response with dose was found to be better than 99%. The independence of signal on dose rate was demonstrated by means of stability of pixel response over the range of dose rates allowed by the radiation source (80-400 MU/min). Finally, excellent long-term stability in pixel response, extending over a 2 month period, was observed. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)02308-1].
引用
收藏
页码:1530 / 1541
页数:12
相关论文
共 31 条
[1]   RADIATION RESPONSE OF AMORPHOUS-SILICON IMAGING ARRAYS AT DIAGNOSTIC ENERGIES [J].
ANTONUK, LE ;
SIEWERDSEN, JH ;
YORKSTON, J ;
HUANG, W .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1500-1505
[2]   Megavoltage imaging with a large-area, flat-panel, amorphous silicon imager [J].
Antonuk, LE ;
Yorkston, J ;
Huang, WD ;
Sandler, H ;
Siewerdsen, JH ;
ElMohri, Y .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (03) :661-672
[3]   Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation [J].
Antonuk, LE ;
ElMohri, Y ;
Siewerdsen, JH ;
Yorkston, J ;
Huang, W ;
Scarpine, VE ;
Street, RA .
MEDICAL PHYSICS, 1997, 24 (01) :51-70
[4]   Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging [J].
Antonuk, LE ;
El-Mohri, Y ;
Huang, WD ;
Jee, KW ;
Siewerdsen, JH ;
Maolinbay, M ;
Scarpine, VE ;
Sandler, H ;
Yorkston, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (02) :437-454
[5]   DEMONSTRATION OF MEGAVOLTAGE AND DIAGNOSTIC-X-RAY IMAGING WITH HYDROGENATED AMORPHOUS-SILICON ARRAYS [J].
ANTONUK, LE ;
BOUDRY, J ;
HUANG, WD ;
MCSHAN, DL ;
MORTON, EJ ;
YORKSTON, J ;
LONGO, MJ ;
STREET, RA .
MEDICAL PHYSICS, 1992, 19 (06) :1455-1466
[6]   DEVELOPMENT OF HYDROGENATED AMORPHOUS-SILICON SENSORS FOR HIGH-ENERGY PHOTON RADIOTHERAPY IMAGING [J].
ANTONUK, LE ;
YORKSTON, J ;
BOUDRY, J ;
LONGO, MJ ;
JIMENEZ, J ;
STREET, RA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :165-170
[7]  
ANTONUK LE, 1995, P SOC PHOTO-OPT INS, V2431, P216
[8]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .2. PROPERTIES AND MEASUREMENTS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1901-1913
[9]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1883-1900
[10]   The dose response relationship of a liquid-filled electronic portal imaging device [J].
Boellaard, R ;
vanHerk, M ;
Mijnheer, BJ .
MEDICAL PHYSICS, 1996, 23 (09) :1601-1611