Tissue oxygen saturation during colorectal surgery measured by near-infrared spectroscopy: Pilot study to predict anastomotic complications

被引:62
作者
Hirano, Y
Omura, K
Tatsuzawa, Y
Shimizu, J
Kawaura, Y
Watanabe, G
机构
[1] Saiseikai Kanazawa Hosp, Dept Surg, Kanazawa, Ishikawa 9200353, Japan
[2] Kanazawa Univ, Sch Med, Dept Gen & Cardiothorac Surg, Kanazawa, Ishikawa 9208641, Japan
关键词
D O I
10.1007/s00268-005-0271-y
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction: We investigated the relation between tissue oxygen saturation measured by near-infrared spectroscopy (NIRS) and anastomotic complications associated with colorectal surgery. Methods: A series of 20 patients with colorectal cancer underwent radical surgery with enteric anastomosis. Measurements of tissue oxygen saturation (StO(2)) were performed at both the proximal and distal portions of the anastomotic site; in cases of anterior resection, we measured StO(2) only in the proximal portion. Reults: Two anastomotic complications (one leakage, one stenosis) occurred in the 20 cases. The StO(2) in patients with anastomotic complications was 58.0%, and that in patients without complications was 71.0%. Altogether, 18 patients had StO(2) values > 66%, and none of them had anastomotic complications. In contrast, 2 patients had StO(2) values < 60%, and both had anastomotic complications. Conclusions The StO(2) of the anastomotic site can be safely and reliably measured by NIRS during colorectal surgery. Low StO(2) on both sides of the anastomosis may indicate an increased risk of anastomotic complications. Further study is needed to determine the cutoff value for StO(2) required to prevent serious complications.
引用
收藏
页码:457 / 461
页数:5
相关论文
共 22 条
[1]
Betzler M, 1998, CHIRURG, V69, P1, DOI 10.1007/s001040050367
[2]
Boyle NH, 1998, BRIT J SURG, V85, P1407
[3]
Changes in tissue oxygenation of the porcine liver measured by near-infrared spectroscopy [J].
El-Desoky, AEH ;
Seifalian, A ;
Cope, M ;
Delpy, D ;
Davidson, B .
LIVER TRANSPLANTATION AND SURGERY, 1999, 5 (03) :219-226
[4]
FREEMAN DE, 1988, AM J VET RES, V49, P895
[5]
NEAR-INFRARED MONITORING OF HUMAN SKELETAL-MUSCLE OXYGENATION DURING FOREARM ISCHEMIA [J].
HAMPSON, NB ;
PIANTADOSI, CA .
JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (06) :2449-2457
[6]
Near-infrared spectroscopy for assessment of tissue oxygen saturation of transplanted jejunal autografts in cervical esophageal reconstruction [J].
Hirano, Y ;
Omura, K ;
Yoshiba, H ;
Ohta, N ;
Hiranuma, C ;
Nitta, K ;
Nishida, Y ;
Watanabe, G .
SURGERY TODAY, 2005, 35 (01) :67-72
[7]
HORVATH KA, 1994, J THORAC CARDIOV SUR, V107, P220
[8]
HULTEN L, 1977, GASTROENTEROLOGY, V72, P388
[9]
HULTEN L, 1976, GASTROENTEROLOGY, V70, P45
[10]
Clinical significance of tissue blood flow during esophagectomy by laser Doppler flowmetry [J].
Ikeda, Y ;
Niimi, M ;
Kan, S ;
Shatari, T ;
Takami, H ;
Kodaira, S .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 122 (06) :1101-1106