Precision and discriminatory ability of calcaneal bone assessment technologies

被引:158
作者
Greenspan, SL
Bouxsein, ML
Melton, ME
Kolodny, AH
Clair, JH
Delucca, PT
Stek, M
Faulkner, KG
Orwoll, ES
机构
[1] MERCK & CO INC, W POINT, PA 19486 USA
[2] BETH ISRAEL DEACONESS MED CTR, OSTEOPOROSIS PREVENT & TREATMENT CTR, BOSTON, MA USA
[3] BETH ISRAEL DEACONESS MED CTR, ORTHOPAED BIOMECH LAB, BOSTON, MA USA
[4] OREGON OSTEOPOROSIS CTR, PORTLAND, OR USA
[5] PORTLAND VA MED CTR, BONE & MINERAL RES UNIT, PORTLAND, OR USA
[6] OREGON HLTH SCI UNIV, PORTLAND, OR 97201 USA
关键词
D O I
10.1359/jbmr.1997.12.8.1303
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To determine if measuring skeletal status at the calcaneus is a potentially valuable technique for diagnosing osteoporosis, we examined five calcaneal assessment techniques in 53 young normal women and 108 postmenopausal women with osteoporosis and compared these measurements to dual-energy X-ray absorptiometry (DEXA)) at the calcaneus, hip, and spine. The five instruments, including single-energy X-ray absorptiometry (SEXA) and four quantitative ultrasound (QUS) instruments, were evaluated for precision, ability to discriminate osteoporotic from young normal subjects, and correlation to the other instruments, The coefficient of variation (%CV) for instrument, positioning, interobserver, and short-term precision of the five calcaneal instruments ranged from 1.34-7.76%, 1.63-7.00%, 1.84-9.44%, and 1.99-7.04%, respectively, The %CVs for positioning, interobserver, and short-term precision were similar for calcaneal DEXA, calcaneal SEXA, and stiffness (as measured by Achilles), The %CVs for instrument precision were similar between calcaneal DEXA and SEXA. The ability of the five calcaneal instruments to discriminate osteoporotic from young normal subjects was similar based on the analysis of area under the receiver operating characteristic curves (range 0.88-0.93) and equivalent to DEXA of the calcaneus and hip (0.88-0.93), The correlations between the measurements of five calcaneal instruments were strong (0.80 less than or equal to r less than or equal to 0.91,p < 0.001), These data suggest that although the precision is variable? the calcaneal QUS and SEXA instruments can discriminate between osteoporotic patients and young normal controls and appear to be a useful technique for assessment of osteoporosis.
引用
收藏
页码:1303 / 1313
页数:11
相关论文
共 41 条
[1]   ULTRASOUND ATTENUATION OF THE CALCANEUS - A SENSITIVE AND SPECIFIC DISCRIMINATOR OF OSTEOPENIA IN POSTMENOPAUSAL WOMEN [J].
AGREN, M ;
KARELLAS, A ;
LEAHEY, D ;
MARKS, S ;
BARAN, D .
CALCIFIED TISSUE INTERNATIONAL, 1991, 48 (04) :240-244
[2]   ULTRASOUND ATTENUATION OF THE OS CALCIS IN WOMEN WITH OSTEOPOROSIS AND HIP-FRACTURES [J].
BARAN, DT ;
KELLY, AM ;
KARELLAS, A ;
GIONET, M ;
PRICE, M ;
LEAHEY, D ;
STEUTERMAN, S ;
MCSHERRY, B ;
ROCHE, J .
CALCIFIED TISSUE INTERNATIONAL, 1988, 43 (03) :138-142
[3]  
BAUER DC, 1995, J BONE MINER RES, V10, P353
[4]  
BLACK DM, 1992, J BONE MINER RES, V7, P633
[5]  
BONNELLI S, 1996, OSTEOPOROSIS INT, V6, P303
[6]   ULTRASOUND AND DUAL X-RAY ABSORPTIOMETRY MEASUREMENT OF THE CALCANEUS - INFLUENCE OF REGION OF INTEREST LOCATION [J].
BROOKEWAVELL, K ;
JONES, PRM ;
PYE, DW .
CALCIFIED TISSUE INTERNATIONAL, 1995, 57 (01) :20-24
[7]   RISK-FACTORS FOR HIP FRACTURE IN WHITE WOMEN [J].
CUMMINGS, SR ;
NEVITT, MC ;
BROWNER, WS ;
STONE, K ;
FOX, KM ;
ENSRUD, KE ;
CAULEY, JC ;
BLACK, D ;
VOGT, TM .
NEW ENGLAND JOURNAL OF MEDICINE, 1995, 332 (12) :767-773
[8]   GERONTOLOGY AND GERIATRICS IN MEDICAL-EDUCATION [J].
DANS, PE ;
KERR, MR .
NEW ENGLAND JOURNAL OF MEDICINE, 1979, 300 (05) :228-232
[9]   QUANTITATIVE ULTRASOUND OF THE HEEL - CORRELATION WITH DENSITOMETRIC MEASUREMENTS AT DIFFERENT SKELETAL SITES [J].
FAULKNER, KG ;
MCCLUNG, MR ;
COLEMAN, LJ ;
KINGSTONSANDAHL, E .
OSTEOPOROSIS INTERNATIONAL, 1994, 4 (01) :42-47
[10]  
GENANT HK, 1995, P 11 INT BON DENS WO