Comparison of optical, electrical, and centrifugation techniques for haematocrit monitoring of dialysed patients

被引:17
作者
Jaffrin, MY [1 ]
Fournier, C [1 ]
机构
[1] Univ Technol Compiegne, Dept Biol Engn, CNRS, UMR 6600, F-60205 Compiegne, France
关键词
haematocrit monitoring; haemodialysis; impedance; blood conductivity;
D O I
10.1007/BF02513326
中图分类号
TP39 [计算机的应用];
学科分类号
081203 [计算机应用技术]; 0835 [软件工程];
摘要
Haematocrits were measured as a function of ultrafiltration in a simulated haemodialysis circuit using bovine blood (plasma conductivity 12 mS cm(-1)) and hypotonic (8. 6 mS cm(-1)) or hypertonic (16 mS cm(-1)) dialysates as well as in the absence of dialysate. A comparison was made between measurements by light absorption due to haemoglobin, by impedance in the blood line at 5 kHz using Hanai's model of blood conductivity, by conductivity measurements of blood samples at 1.2 kHz using a conductimeter, by centrifugation of blood samples and by calculations using fluid conservation. The validity of Hanai's model was verified to be satisfactory by direct blood and plasma conductivity measurements. In the absence of ionic transfer the impedance device underestimated the haematocrit by 5 to 7%. This underestimation reached 18% in the case of hypertonic dialysate, but this effect can be minimised if the haematocrit necessary for calibration is measured by centrifugation after 15 min of dialysate circulation when ionic balance is achieved. It was found that the optical method monitors haemoglobin concentration rather than red cell volume changes and is not affected by osmotic red cell swelling in the case of hypotonic dialysate. It can be concluded that the light absorption technique is both more accurate and more convenient to use than impedance.
引用
收藏
页码:433 / 439
页数:7
相关论文
共 12 条
[1]
de Vries J P, 1992, ASAIO J, V38, pM181, DOI 10.1097/00002480-199207000-00015
[2]
de Vries P M, 1988, ASAIO Trans, V34, P623
[3]
MEASUREMENT OF TRANS-CELLULAR FLUID SHIFT DURING HEMODIALYSIS .2. INVITRO AND CLINICAL-EVALUATION [J].
DEVRIES, PMJM ;
MEIJER, JH ;
VLAANDEREN, K ;
VISSER, V ;
OE, PL ;
DONKER, AJM ;
SCHNEIDER, H .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1989, 27 (02) :152-158
[4]
SPECIFIC RESISTANCE OF BLOOD AT BODY-TEMPERATURE .2. [J].
GEDDES, LA ;
KIDDER, H .
MEDICAL & BIOLOGICAL ENGINEERING, 1976, 14 (02) :180-185
[5]
Hanai T., 1968, EMULSION SCI
[6]
Extra- and intracellular volume monitoring by impedance during haemodialysis using Cole-Cole extrapolation [J].
Jaffrin, MY ;
Maasrani, M ;
LeGourrier, A ;
Boudailliez, B .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (03) :266-270
[7]
KANAI H, 1976, 11 INT C MED BIOL EN
[8]
Continuous measurements by impedance of haematocrit and plasma volume variations during dialysis [J].
Maasrani, M ;
Jaffrin, MY ;
Boudailliez, B .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (03) :167-171
[9]
NEWLY DEVELOPED CIRCULATING BLOOD VOLUME-MONITORING SYSTEM AND ITS CLINICAL-APPLICATION FOR MEASURING CHANGES IN BLOOD-VOLUME DURING HEMOFILTRATION [J].
MAEDA, K ;
SHINZATO, T ;
YOSHIDA, F ;
TSURUTA, Y ;
USUDA, M ;
YAMADA, K ;
ISHIHARA, T ;
INAGAKI, F ;
IGARASHI, I ;
KITANO, T .
ARTIFICIAL ORGANS, 1986, 10 (06) :452-459
[10]
CHANGES IN BLOOD RESISTIVITY WITH HEMATOCRIT AND TEMPERATURE [J].
MOHAPATRA, SN ;
HILL, DW .
EUROPEAN JOURNAL OF INTENSIVE CARE MEDICINE, 1975, 1 (04) :153-162