A concentric-ring equivalent phased array method to model fields of large axisymmetric ultrasound transducers

被引:7
作者
Fan, XB [1 ]
Moros, EG
Straube, WL
机构
[1] Univ Chicago, Med Ctr, Dept Radiol, Chicago, IL 60637 USA
[2] Washington Univ, Sch Med, Ctr Radiat Oncol, St Louis, MO 63108 USA
关键词
D O I
10.1109/58.775647
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
An equivalent concentric-ring ultrasound phased array method was developed to estimate ultrasonic continuous wave fields generated by axisymmetric single-source transducers, The method models a given source as a concentric-ring phased array by mathematically segmenting it into many rings and subsequently finding the amplitude and phase for each ring that produces an acoustic field similar to the held of the single-source transducer. The excitation source of each ring was calculated using an inverse technique based on complex pressure measurements along a radial line close to the source. The predicting abilities of the method are evaluated by comparing measured and estimated ultrasound fields for six different transducers. The results show that the concentric-ring equivalent phased array method (CREPAM) is able to estimate quantitatively the ultrasound fields generated by large axisymmetric planar and focused transducers.
引用
收藏
页码:830 / 841
页数:12
相关论文
共 35 条
[1]   TRANSIENT FIELDS OF CONCAVE ANNULAR ARRAYS [J].
ARDITI, M ;
FOSTER, FS ;
HUNT, JW .
ULTRASONIC IMAGING, 1981, 3 (01) :37-61
[2]   CONCENTRIC-RING AND SECTOR-VORTEX PHASED-ARRAY APPLICATORS FOR ULTRASOUND HYPERTHERMIA [J].
CAIN, CA ;
UMEMURA, S .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (05) :542-551
[3]  
CHAPELON JY, 1993, ULTR S, V93, P1211
[4]   PIEZOELECTRIC POLYMER PROBE FOR ULTRASONIC APPLICATIONS [J].
DEREGGI, AS ;
ROTH, SC ;
KENNEY, JM ;
EDELMAN, S ;
HARRIS, GR .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1981, 69 (03) :853-859
[5]  
DOHUU JP, 1981, ULTR S P IEEE 81, P705
[7]   A SPHERICAL-SECTION ULTRASOUND PHASED-ARRAY APPLICATOR FOR DEEP LOCALIZED HYPERTHERMIA [J].
EBBINI, ES ;
CAIN, CA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (07) :634-643
[8]   Acoustic field prediction for a single planar continuous-wave source using an equivalent phased array method [J].
Fan, XB ;
Mores, EG ;
Straube, WL .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 102 (05) :2734-2741
[9]   A parametric study of the concentric-ring transducer design for MRI guided ultrasound surgery [J].
Fjield, T ;
Fan, XB ;
Hynynen, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1996, 100 (02) :1220-1230
[10]   SOUND FIELD OF A RECTANGULAR PISTON [J].
FREEDMAN, A .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1960, 32 (02) :197-209