Cleveland clinic CorAide blood pump circulatory support without anticoagulation

被引:9
作者
Ochiai, Y [1 ]
Golding, LAR [1 ]
Massiello, AL [1 ]
Medvedev, AL [1 ]
Horvath, DJ [1 ]
Gerhart, RL [1 ]
Chen, JF [1 ]
Krogulecki, AY [1 ]
Takagaki, M [1 ]
Doi, K [1 ]
Howard, MW [1 ]
Fukamachi, K [1 ]
机构
[1] Cleveland Clin Fdn, Dept Biomed Engn ND20, Lerner Res Inst, Cleveland, OH 44195 USA
关键词
D O I
10.1097/00002480-200205000-00009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Cleveland Clinic CorAide left ventricular assist system consists of a permanently implantable centrifugal pump in which the rotating assembly is completely suspended and noncontacting. A series of chronic animal in vivo studies were conducted to evaluate the biologic effects of CorAide circulatory support without the use of anticoagulation therapy. The CorAide pump was implanted in six calves (five calves for 21 to 32 days and one calf for 95 days). The first five calves received intravenous heparin during the early postoperative periods (2-7 days). Heparin administration was then discontinued and no other anticoagulant drugs were used for the duration of the experiments. The last calf did not receive any anticoagulant except for a bolus dose of heparin (200 U/kg) during surgery. Hemodynamics were stable in all six calves, with a mean pump flow of 5.6 +/- 1.2 L/min and mean arterial pressure of 100 +/- 4 mm Hg. The blood pump surfaces were clean of thrombus in all six calves. Significant findings at autopsy were limited to one case of renal infarction. There was no incidence of mechanical failure, bleeding, or device infection. The CorAide pump can be safely run with minimal or no anticoagulant therapy.
引用
收藏
页码:249 / 252
页数:4
相关论文
共 15 条
[1]   An anatomically compatible, wearless, reliable, and nonthrombogenic centrifugal blood pump [J].
Boyce, SW ;
Crevensten, G ;
Fine, RB .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S190-S190
[2]   Research and development of an implantable, axial-flow left ventricular assist device: The Jarvik 2000 heart [J].
Frazier, OH ;
Myers, TJ ;
Jarvik, RK ;
Westaby, S ;
Pigott, DW ;
Gregoric, ID ;
Khan, T ;
Tamez, DW ;
Conger, JL ;
Macris, MP .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S125-S132
[3]   HeartMate II left ventricular assist system: From concept to first clinical use [J].
Griffith, BP ;
Kormos, RL ;
Borovetz, HS ;
Litwak, K ;
Antaki, JF ;
Poirier, VL ;
Butler, KC .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S116-S120
[4]   The CorAide blood pump [J].
Horvath, DJ ;
Golding, LAR ;
Massiello, A ;
Medvedev, AL ;
Gerhart, RL ;
Ochiai, Y ;
Krogulecki, A ;
Fukamachi, K .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S191-S191
[5]   Factors influencing HLA sensitization implantable LVAD recipients [J].
Massad, MG ;
Cook, DJ ;
Schmitt, SK ;
Smedira, NG ;
McCarthy, JF ;
Vargo, RL ;
McCarthy, PM .
ANNALS OF THORACIC SURGERY, 1997, 64 (04) :1120-1125
[6]   Clinical experience with the MicroMed DeBakey ventricular assist device [J].
Noon, GP ;
Morley, DL ;
Irwin, S ;
Abdelsayed, SV ;
Benkowski, RJ ;
Lynch, BE .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S133-S138
[7]   In vivo hemodynamic performance of the Cleveland Clinic CorAide blood pump in calves [J].
Ochiai, Y ;
Golding, LAR ;
Massiello, AL ;
Medvedev, AL ;
Gerhart, RL ;
Chen, JF ;
Takagaki, M ;
Fukamachi, K .
ANNALS OF THORACIC SURGERY, 2001, 72 (03) :747-752
[8]   Improved outcomes with an implantable left ventricular assist system: A multicenter study [J].
Portner, PM ;
Jansen, PGM ;
Oyer, PE ;
Wheeldon, DR ;
Ramasamy, N .
ANNALS OF THORACIC SURGERY, 2001, 71 (01) :205-209
[9]   Reliable long-term non-pulsatile circulatory support without anticoagulation [J].
Saito, S ;
Westaby, S ;
Piggott, D ;
Katsumata, T ;
Dudnikov, S ;
Robson, D ;
Catarino, P ;
Nojiri, C .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2001, 19 (05) :678-683
[10]   Mechanical support - Benefits and risks [J].
Scheld, HH .
THORACIC AND CARDIOVASCULAR SURGEON, 1997, 45 (01) :1-5