Patient motion during brain SPECT studies can degrade resolution and introduce distortion. We have developed a correction method which incorporates a motion tracking system to monitor the position and orientation of the patient's head during acquisition. Correction is achieved by spatially repositioning projections according to measured head movements and reconstructing these projections with a fully three-dimensional (3D) algorithm. The method has been evaluated in SPECT studies of the Hoffman 3D brain phantom performed on a triple head camera with fan beam collimation. Movements were applied to the phantom and recorded by a head tracker during SPECT acquisition. Fully 3D reconstruction was performed using the motion data provided by the tracker. Correction accuracy was assessed by comparing the corrected and uncorrected studies with a motion free study, visually and by calculating mean squared error (MSE). In all studies, motion correction reduced distortion and improved MSE by a factor of 2 or more. We conclude that this method can compensate for head motion under clinical SPECT imaging conditions.
引用
收藏
页码:667 / 672
页数:6
相关论文
共 13 条
[1]  
Chen Q., 1993, J NUCL MED TECHNOL, V21, P198
[2]  
DAUBEWITHERSPOO.ME, 1990, J NUCL MED, V31, P186
[3]   USE OF 3D RECONSTRUCTION TO CORRECT FOR PATIENT MOTION IN SPECT [J].
FULTON, RR ;
HUTTON, BF ;
BRAUN, M ;
ARDEKANI, B ;
LARKIN, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (03) :563-574
[4]  
GREEN MV, 1994, J NUCL MED, V35, P1538
[5]   3-D PHANTOM TO SIMULATE CEREBRAL BLOOD-FLOW AND METABOLIC IMAGES FOR PET [J].
HOFFMAN, EJ ;
CUTLER, PD ;
DIGBY, WM ;
MAZZIOTTA, JC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :616-620
[6]   CLOSED-FORM SOLUTION OF ABSOLUTE ORIENTATION USING UNIT QUATERNIONS [J].
HORN, BKP .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (04) :629-642
[7]   ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA [J].
HUDSON, HM ;
LARKIN, RS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :601-609
[8]   Use of forward projection to correct patient motion during SPECT imaging [J].
Lee, KJ ;
Barber, DC .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (01) :171-187
[9]   A filtered backprojection algorithm for axial head motion correction in fan-beam SPECT [J].
Li, JY ;
Jaszczak, RJ ;
Coleman, RE .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12) :2053-2063
[10]   Compensation methods for head motion detected during PET imaging [J].
Menke, M ;
Atkins, MS ;
Buckley, KR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (01) :310-317