Nasal bridle revisited: An improvement in the technique to prevent unintentional removal of small-bore nasoenteric feeding tubes

被引:30
作者
Popovich, MJ [1 ]
Lockrem, JD [1 ]
Zivot, JB [1 ]
机构
[1] UNIV COLORADO,HLTH SCI CTR,DEPT ANESTHESIOL,DENVER,CO
关键词
enteral nutrition; tube feeding; complications; oropharynx; intubation; nasoenteric; critical illness;
D O I
10.1097/00003246-199603000-00011
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To demonstrate the efficacy and safety of an up-dated version of the nasal ''bridle,'' which is used to prevent the accidental removal of small-bore nasoenteric feeding tubes. Design: A descriptive study. Setting: Surgical intensive care unit in a tertiary care hospital. Patients: Twenty-six critically ill patients without nasotracheal tubes or facial trauma or fractures who received enteral nutrition and either had removed or were at risk for removing their properly positioned nasoenteric feeding tubes. Interventions: A length of one-eighth inch (3.2 mm) umbilical tape is looped around the nasal septum and vomer by serially attaching the ends of the umbilical tape to a suction catheter, passing the catheter through the nostrils into the oropharynx, and retrieving the ends from the oropharynx, The properly positioned umbilical tape loops into one nostril, around the vomer, and out the other nostril, The feeding tube is then anchored to the umbilical tape with a central venous catheter fastener clamp. Measurements and Main Results: Communicative patients denied discomfort, and there were no episodes of bleeding, infection, sinusitis, or nasal septal trauma caused by the umbilical tape bridle, Five patients had the bridle in place >30 days, There were only two cases in which the bridle failed to prevent removal of a feeding tube, One of these cases occurred because the fastener clamp anchor failed, but this patient had had the same bridle and feeding tube for 170 consecutive days. Conclusions: An umbilical tape bridle with a central venous catheter fastener clamp anchor is a safe and effective method to prevent the accidental removal of nasoenteric feeding tubes in critically ill patients, We recommend its use in confused or uncooperative patients, or when the risk of unintentional feeding tube removal is high.
引用
收藏
页码:429 / 431
页数:3
相关论文
共 11 条
[1]  
Armstrong C, 1980, JPEN-PARENTER ENTER, V3, P603, DOI [10.1016/j.ijnurstu.2013.12.002, DOI 10.1016/J.IJNURSTU.2013.12.002]
[2]  
Barrocas A., 1982, NUTR SUPPORT SERV, V2, P8
[3]  
Dark DS, 1993, J INTENSIVE CARE MED, V8, P16
[4]  
Keohane P P, 1983, Clin Nutr, V2, P85, DOI 10.1016/0261-5614(83)90038-9
[5]  
KUDSK KA, 1994, NUTR CRITICAL CARE, P331
[6]  
Levensen R., 1985, NUTR SUPP SERV, V5, P40
[7]   DO WEIGHTED NASOENTERIC FEEDING TUBES FACILITATE DUODENAL INTUBATIONS [J].
LEVENSON, R ;
TURNER, WW ;
DYSON, A ;
ZIKE, L ;
REISCH, J .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1988, 12 (02) :135-137
[8]   A NEW NASAL BRIDLE FOR SECURING NASOENTERAL FEEDING TUBES [J].
MEER, JA .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1989, 13 (03) :331-334
[9]   INADVERTENT DISLODGMENT OF NASOENTERAL FEEDING TUBES - INCIDENCE AND PREVENTION [J].
MEER, JA .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1987, 11 (02) :187-189
[10]   ENTERAL FEEDING TUBES - PLACEMENT BY USING FLUOROSCOPY AND ENDOSCOPY [J].
OTT, DJ ;
MATTOX, HE ;
GELFAND, DW ;
CHEN, MYM ;
WU, WC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 157 (04) :769-771