Resonance frequency assessment of dental implant stability with various bone qualities: a numerical approach

被引:141
作者
Huang, HM
Lee, SY
Yeh, CY
Lin, CT
机构
[1] Taipei Med Univ, Grad Inst Oral Rehabil Sci, Taipei, Taiwan
[2] Taipei Med Univ, Sch Med Technol, Taipei, Taiwan
[3] Taipei Med Univ, Dept Dent, Wan Fang Hosp, Taipei, Taiwan
[4] Taipei Med Univ, Sch Med, Taipei, Taiwan
关键词
resonance frequency; dental implant; stability; finite element; biomechanics;
D O I
10.1034/j.1600-0501.2002.130108.x
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
Resonance frequency analysis (RFA) has been used by several investigators to assess the boundary conditions of dental implants. The goal of the current study was to determine the vibrating behavior of a dental implant under various surrounding bone conditions. A 3D finite element (FE) model of a cylinder-type titanium implant was developed. In this model, the implant was embedded into a cubic section of bone. The model was first validated using a series of modal testing experiments. The effects of bony conditions on the resonance frequencies of the implant were computed with different bone types and bone densities. Our results show that the resonance frequency of the implant with type III surrounding bone decreased linearly (r=-0.996, P<0.01) from 17.9 kHz (without loss in bone density) to 0.6 kHz (90% loss in bone density) when the bone densities were decreased. On the other hand, without bone loss, the highest resonance frequency value (36.1 kHz) was found when the implant was placed into type I surrounding bone. In contrast, the resonance frequency of the implant with type IV bone quality was found to be 9.9 kHz, which is almost four-fold less than that found in the type I model. These results suggest that RFA could serve as a non-invasive diagnostic tool for detecting the stability of dental implants during the healing stages and in subsequent routine follow-up care after treatment.
引用
收藏
页码:65 / 74
页数:10
相关论文
共 49 条
[1]
CLINICAL-RESULTS OF OSSEOINTEGRATED IMPLANTS SUPPORTING FIXED PROSTHESES IN EDENTULOUS JAWS [J].
ADELL, R .
JOURNAL OF PROSTHETIC DENTISTRY, 1983, 50 (02) :251-254
[2]
STATE-OF-THE-ART IN ORAL IMPLANTS [J].
ALBREKTSSON, T ;
SENNERBY, L .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1991, 18 (06) :474-481
[3]
COMPRESSIVE BEHAVIOR OF BONE AS A 2-PHASE POROUS STRUCTURE [J].
CARTER, DR ;
HAYES, WC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (07) :954-962
[4]
The design of a vibration transducer to monitor the integrity of dental implants [J].
Cawley, P ;
Pavlakovic, B ;
Alleyne, DN ;
George, R ;
Back, T ;
Meredith, N .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1998, 212 (H4) :265-272
[5]
CHRISTENSEN AB, 1986, J BIOMECH, V19, P53, DOI 10.1016/0021-9290(86)90108-9
[6]
ASSESSMENT OF TIBIAL STIFFNESS BY VIBRATION TESTING INSITU .2. INFLUENCE OF SOFT-TISSUES, JOINTS AND FIBULA [J].
CORNELISSEN, P ;
CORNELISSEN, M ;
VANDERPERRE, G ;
CHRISTENSEN, AB ;
AMMITZBOLL, F ;
DYRBYE, C .
JOURNAL OF BIOMECHANICS, 1986, 19 (07) :551-561
[7]
Elias J J, 1996, Int J Oral Maxillofac Implants, V11, P728
[8]
EWINS DJ, 1986, MODAL TESTING THEORY, P88
[9]
Stability measurements of one-stage Branemark implants during healing in mandibles -: A clinical resonance frequency analysis study [J].
Friberg, B ;
Sennerby, L ;
Linden, B ;
Gröndahl, K ;
Lekholm, U .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1999, 28 (04) :266-272
[10]
NATURAL FREQUENCY-ANALYSIS OF A HUMAN TIBIA [J].
HIGHT, TK ;
PIZIALI, RL ;
NAGEL, DA .
JOURNAL OF BIOMECHANICS, 1980, 13 (02) :139-147