Diagnostic accuracy of a new software for complex fractionated electrograms identification in patients with persistent and permanent atrial fibrillation

被引:23
作者
Calo, Leonardo [1 ]
De Ruvo, Ermenegildo [1 ]
Sciarra, Luigi [1 ]
Gricia, Roberto [2 ]
Navone, Giovanna [1 ]
De Luca, Lucia [1 ]
Nuccio, Francesca [1 ]
Sette, Antonella [1 ]
Pristipino, Cristian [3 ]
Dulio, Alessandro [2 ]
Gaita, Fiorenzo [4 ]
Lioy, Ernesto [1 ]
机构
[1] Policlin Casilino, Div Cardiol, ASL Roma B, I-00047 Rome, Italy
[2] Johnson & Johnson Med, Biosense Webster Italy, Milan, Italy
[3] San Filippo Neri Hosp, Dept Cardiovasc Dis, Rome, Italy
[4] Osped Civile, Div Cardiol, Asti, Italy
关键词
atrial fibrillation; mapping; catheter ablation; complex fractionated atrial electrograms;
D O I
10.1111/j.1540-8167.2008.01219.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
CARTO Mapping of CFAEs. Introduction: The elimination of complex fractionated atrial electrograms (CFAEs) has been proposed as a potential target for guiding successful AF substrate ablation. The possibility to efficiently map the atria and rapidly identify CFAEs sites is necessary, before the CFAEs ablation becomes a routine approach. The aims of this study, conducted in patients with persistent and permanent atrial fibrillation (AF), were to analyze by CARTO mapping in the right (RA) and in the left atrium (LA) during AF: (1) the diagnostic accuracy of a new software for CFAEs analysis, (2) the spatial distribution of CFAEs, (3) the regional beat to beat AF intervals (FF). Methods and Results: Twenty-five consecutive patients (four women, 58.8 +/- 11.4 years) undergoing radiofrequency catheter ablation for persistent and permanent AF were enrolled in the study. The CFAE software showed a high sensitivity (90%) and specificity (91%) in the identification of CFAEs, using a specific setting of parameters. The LA had a significantly higher prevalence of CFAEs as compared with the RA (30.5% vs 20.3%, P = 0.016). The CFAEs were mostly present in the septum and in the area of coronary sinus ostium (CS os). The FF intervals were significantly shorter in the LA than in the RA (P < 0.01). Conclusion: CARTO system has a high diagnostic accuracy in the identification of CFAEs. Atrial electrical activity (CFAEs, mean FF intervals) during AF showed a significant spatial inhomogeneity.
引用
收藏
页码:1024 / 1030
页数:7
相关论文
共 30 条
[1]   Feasibility of a left atrial electrical disconnection procedure for atrial fibrillation using transcatheter radiofrequency ablation [J].
Betts, TR ;
Roberts, PR ;
Morgan, JM .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (11) :1278-1283
[2]   Left atrial ablation versus biatrial ablation for persistent and permanent atrial fibrillation -: A prospective and randomized study [J].
Calo, Leonardo ;
Lamberti, Filippo ;
Loricchio, Maria Luisa ;
De Ruvo, Ermenegildo ;
Colivicchi, Furio ;
Bianconi, Leopoldo ;
Pandozi, Claudio ;
Santini, Massimo .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (12) :2504-2512
[3]   Myocyte changes and their left atrial distribution in patients with chronic atrial fibrillation related to mitrat valve disease [J].
Corradi, D ;
Callegari, S ;
Benussi, S ;
Maestri, R ;
Pastori, P ;
Nascimbene, S ;
Bosio, S ;
Dorigo, E ;
Grassani, C ;
Rusconi, R ;
Vettori, MV ;
Alinovi, R ;
Astorri, E ;
Pappone, C ;
Alfieri, O .
HUMAN PATHOLOGY, 2005, 36 (10) :1080-1089
[4]   An 8 1/2-year clinical experience with surgery for atrial fibrillation [J].
Cox, JL ;
Schuessler, RB ;
Lappas, DG ;
Boineau, JP .
ANNALS OF SURGERY, 1996, 224 (03) :267-273
[5]   Different patterns of atrial activation in idiopathic atrial fibrillation:: Simultaneous multisite atrial mapping in patients with paroxysmal and chronic atrial fibrillation [J].
Gaita, F ;
Calò, L ;
Riccardi, R ;
Garberoglio, L ;
Scaglione, M ;
Licciardello, G ;
Coda, L ;
Di Donna, P ;
Bocchiardo, M ;
Caponi, D ;
Antolini, R ;
Orzan, F ;
Trevi, G .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) :534-541
[6]   Atrial mapping and radiofrequency catheter ablation inpatients with idiopathic atrial fibrillation -: Electrophysiological findings and ablation results [J].
Gaita, F ;
Riccardi, R ;
Calò, L ;
Scaglione, M ;
Garberoglio, L ;
Antolini, R ;
Kirchner, M ;
Lamberti, F ;
Richiardi, E .
CIRCULATION, 1998, 97 (21) :2136-2145
[7]   Catheter ablation of long-lasting persistent atrial fibrillation:: Critical structures for termination [J].
Haïssaguerre, M ;
Sanders, P ;
Hocini, M ;
Takahashi, Y ;
Rotter, M ;
Sacher, F ;
Rostock, T ;
Hsu, LF ;
Bordachar, P ;
Reuter, S ;
Roudaut, R ;
Clémenty, J ;
Jaïs, P .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (11) :1125-1137
[8]   Ganglionated plexi modulate extrinsic cardiac autonomic nerve input - Effects on sinus rate, atrioventricular conduction, refractoriness, and inducibility of atrial fibrillation [J].
Hou, Yinglong ;
Scherlag, Benjamin J. ;
Lin, Jiaxiong ;
Zhang, Ying ;
Lu, Zhibing ;
Truong, Kim ;
Patterson, Eugene ;
Lazzara, Ralph ;
Jackman, Warren M. ;
Po, Sunny S. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (01) :61-68
[9]   Regional disparities of endocardial atrial activation in paroxysmal atrial fibrillation [J].
Jais, P ;
Haissaguerre, M ;
Shah, DC ;
Chouairi, S ;
Clementy, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (11) :1998-2003
[10]   Mother rotors and fibrillatory conduction: a mechanism of atrial fibrillation [J].
Jalife, J ;
Berenfeld, O ;
Mansour, M .
CARDIOVASCULAR RESEARCH, 2002, 54 (02) :204-216