A novel ASIC for readout electronics in semiconductor gamma-ray detection

被引:4
作者
Bonnefoy, JP [1 ]
Lattard, D [1 ]
Mathy, F [1 ]
Martin, JL [1 ]
Poujois, R [1 ]
Rostaing, JP [1 ]
Trystram, P [1 ]
Cretolle, J [1 ]
机构
[1] CEA SACLAY,DAPNIA,SERV ASTROPHYS,F-91191 GIF SUR YVETTE,FRANCE
关键词
D O I
10.1016/S0168-9002(96)00382-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A low-power low-noise multichannel integrated circuit is implemented in 1.2 mu m CMOS technology. This circuit is designed for space gamma-ray spectrometer using arrays of CdTe radiation detectors in the 20-750 keV range. The space environment needs rad-tolerant specifications for the technology like 20 krad of total integrated dose and consumption limited to 2 mW/channel. Each channel includes a low-noise charge-sensitive preamplifier, fitted band-pass filters, peak amplitude and rise-time measurements; these enable a remote rise-time corrected gamma-ray energy spectral analysis. The circuit includes an associated logic block, dedicated to the internal timing, functional organization, external communication and data transfer.
引用
收藏
页码:342 / 345
页数:4
相关论文
共 3 条
[1]  
DANTONI P, 1994, CEA DAPNIA INTERNAL
[2]   A CdTe gamma-camera for the space observatory INTEGRAL: Software charge-loss corrections [J].
Lebrun, F ;
Leray, JP ;
Laurent, P ;
DeAntoni, P ;
Blondel, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) :414-418
[3]   HIGH-RESOLUTION GAMMA-RAY SPECTROSCOPY WITH CDTE DETECTOR SYSTEMS [J].
RICHTER, M ;
SIFFERT, P .