Carotid-femoral pulse wave velocity assessment using novel cuff-based techniques: comparison with tonometric measurement

被引:86
作者
Butlin, Mark [1 ]
Qasem, Ahmad [1 ]
Battista, Francesca [2 ,3 ]
Bozec, Erwan [4 ]
McEniery, Carmel M. [5 ]
Millet-Amaury, Eugenie [4 ]
Pucci, Giacomo [2 ,3 ]
Wilkinson, Ian B. [5 ]
Schillaci, Giuseppe [2 ,3 ]
Boutouyrie, Pierre [4 ]
Avolio, Alberto P. [1 ]
机构
[1] Macquarie Univ, Australian Sch Adv Med, Sydney, NSW 2109, Australia
[2] Univ Perugia, Terni, Italy
[3] Osped S Maria, Terni, Italy
[4] Univ Paris 05, Hop Europeen Georges Pompidou, AP HP, Dept Pharmacol,INSERM,UMR970, Paris, France
[5] Univ Cambridge, Addenbrookes Hosp, Clin Pharmacol Unit, Cambridge CB2 2QQ, England
关键词
aortic stiffness; arterial compliance; device validation; hemodynamics; measurement; pulse wave velocity; ARTERIAL STIFFNESS; AORTIC STIFFNESS; EVENTS; AGE;
D O I
10.1097/HJH.0b013e328363c789
中图分类号
R6 [外科学];
学科分类号
100210 [外科学];
摘要
Background:Carotid-femoral pulse wave velocity, a predictor of cardiovascular outcome, is conventionally measured using a tonometer sequentially placed upon the carotid and femoral arteries, gated using an electrocardiogram. Leg cuff detection of the femoral pulse removes the need for signal gating, reduces the time required for a single measurement, but gives different pulse wave velocity values to tonometric analysis. A novel algorithm to correct for the transit time and distance related to the additional femoral segment was applied to the cuff-based approach in this study.Method:Eighty-eight individuals were recruited across four centres and carotid-femoral pulse wave velocity measured in triplicate using two operators with both a tonometer-based device and a device using an inflated thigh cuff with and without the use of the novel algorithm. Comparison was made by Bland-Altman and regression analysis.Results:The unadjusted cuff-based approach gave lower pulse wave velocity values than the tonometer-based approach (6.111.27 and 7.02 +/- 1.88m/s, P<0.001). With application of the algorithm, the cuff-based device gave similar pulse wave velocity values (7.04 +/- 1.72m/s) as the tonometer-based approach (P=0.86). Analysis of covariance with age showed a difference between the tonometer and cuff-based methods (P<0.001), with a dependence upon age (P=0.004). The adjusted cuff-based method gave similar results to the tonometer-based method (P=0.94), with no dependence upon age (P=0.46).Conclusion:This study provided validation of a cuff-based assessment of carotid-femoral pulse wave velocity against the universally accepted tonometric method. Adjusting the cuff-based method for the additional femoral segment measured gives results comparable to the tonometer-based method, for which the majority of population data exist to date.
引用
收藏
页码:2237 / 2243
页数:7
相关论文
共 19 条
[1]
Arterial Stiffness, Its Assessment, Prognostic Value, and Implications for Treatment [J].
Adji, Audrey ;
O'Rourke, Michael F. ;
Namasivayam, Mayooran .
AMERICAN JOURNAL OF HYPERTENSION, 2011, 24 (01) :5-17
[2]
EFFECTS OF AGING ON CHANGING ARTERIAL COMPLIANCE AND LEFT-VENTRICULAR LOAD IN A NORTHERN CHINESE URBAN-COMMUNITY [J].
AVOLIO, AP ;
CHEN, SG ;
WANG, RP ;
ZHANG, CL ;
LI, MF ;
OROURKE, MF .
CIRCULATION, 1983, 68 (01) :50-58
[3]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]
Aortic stiffness is an independent predictor of primary coronary events in hypertensive patients - A longitudinal study [J].
Boutouyrie, P ;
Tropeano, AI ;
Asmar, R ;
Gautier, I ;
Benetos, A ;
Lacolley, P ;
Laurent, S .
HYPERTENSION, 2002, 39 (01) :10-15
[5]
Outcome Associations of Carotid-Femoral Pulse Wave Velocity Vary With Different Measurement Methods [J].
Girerd, Nicolas ;
Legedz, Liliana ;
Paget, Vinciane ;
Rabilloud, Muriel ;
Milon, Hugues ;
Bricca, Giampiero ;
Lantelme, Pierre .
AMERICAN JOURNAL OF HYPERTENSION, 2012, 25 (12) :1264-1270
[6]
The Relationship of Age With Regional Aortic Stiffness and Diameter [J].
Hickson, Stacey S. ;
Butlin, Mark ;
Graves, Martin ;
Taviani, Valentina ;
Avolio, Alberto P. ;
McEniery, Carmel M. ;
Wilkinson, Ian B. .
JACC-CARDIOVASCULAR IMAGING, 2010, 3 (12) :1247-1255
[7]
Validity and repeatability of the Vicorder apparatus: a comparison with the SphygmoCor device [J].
Hickson, Stacey S. ;
Butlin, Mark ;
Broad, Jeremy ;
Avolio, Alberto P. ;
Wilkinson, Ian B. ;
McEniery, Carmel M. .
HYPERTENSION RESEARCH, 2009, 32 (12) :1079-1085
[8]
Carotid to femoral pulse wave velocity: a comparison of real travelled aortic path lengths determined by MRI and superficial measurements [J].
Huybrechts, Sofie A. M. ;
Devos, Daniel G. ;
Vermeersch, Sebastian J. ;
Mahieu, Dries ;
Achten, Eric ;
de Backer, Tine L. M. ;
Segers, Patrick ;
van Bortel, Luc M. .
JOURNAL OF HYPERTENSION, 2011, 29 (08) :1577-1582
[9]
Expert consensus document on arterial stiffness: methodological issues and clinical applications [J].
Laurent, Stephane ;
Cockcroft, John ;
Van Bortel, Luc ;
Boutouyrie, Pierre ;
Giannattasio, Cristina ;
Hayoz, Daniel ;
Pannier, Bruno ;
Vlachopoulos, Charalambos ;
Wilkinson, Ian ;
Struijker-Boudier, Harry .
EUROPEAN HEART JOURNAL, 2006, 27 (21) :2588-2605
[10]
Mancia G, 2007, EUR HEART J, V28, P1462, DOI [10.1093/eurheartj/ehm236, 10.1093/eurheartj/eht151]