A model for the harmonisation of test results of different quantitative D-dimer methods

被引:54
作者
Meijer, P
Haverkate, F
Kluft, C
de Moerloose, P
Verbruggen, B
Spannagl, M
机构
[1] ECAT Fdn, NL-2301 CE Leiden, Netherlands
[2] TNO, Prevent & Hlth, Leiden, Netherlands
[3] Univ Hosp Geneva, Geneva, Switzerland
[4] Acad Med Ctr St Radboud, Nijmegen, Netherlands
[5] Univ Hosp Munich, Munich, Germany
[6] Instand, Dusseldorf, Germany
关键词
D-dimer; harmonisation; linear regression; commutability;
D O I
10.1160/TH05-01-0042
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The numerical test results of different D-dimer assays vary widely. Because of the complexity of the analyte of target as well as the variability in specificity of different D-dimer assays, only harmonisation of the test results seems to be feasible. The use of a single conversion factor does not take into account for several methods the lack of commutability between test results and consensus values at different D-dimer levels. This is probably related to the mutually different response of methods to high and low levels. We therefore designed a harmonisation model based on the transformation of a method-specific regression line to a reference regression line. We used the data for the measurement of a set of plasma samples with different D-dimer levels by 353 different laboratories using 7 of the most frequently used quantitative D-dimer methods. For each method we calculated the method-specific consensus value for each sample. The overall median value was also estimated. Per method linear regression was applied throughout the method-specific consensus values using the amount of patient pooled plasma added to the different plasma samples as the independent variable. The line through the overall median values of all 7 methods was used as the reference line. Harmonisation between the methods was obtained by transformation of the method-specific regression line to the reference line. This harmonisation resulted in a reduction of the variability between the method-specific consensus values from about 75% to about 5.5%. Clinical validation of this concept had shown significant improvement of the test result comparability. We conclude that this model is a feasible approach in the harmonisation of D-dimer methods. If the harmonisation procedure is included in the calibration procedure by the manufacturers, customers will automatically obtain harmonised test results.
引用
收藏
页码:567 / 572
页数:6
相关论文
共 17 条
  • [1] de Maat MPM, 2000, SEMIN THROMB HEMOST, V26, P625
  • [2] Dempfle CE, 2001, THROMB HAEMOSTASIS, V85, P671
  • [3] Elias A, 1996, THROMB HAEMOSTASIS, V76, P518
  • [4] Freyburger G, 1998, THROMB HAEMOSTASIS, V79, P32
  • [5] FIBRIN DEGRADATION PRODUCT (FNDP) ASSAYS - ANALYSIS OF STANDARDIZATION ISSUES AND TARGET ANTIGENS IN PLASMA
    GAFFNEY, PJ
    EDGELL, T
    CREIGHTONKEMPSFORD, LJ
    WHEELER, S
    TARELLI, E
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 1995, 90 (01) : 187 - 194
  • [6] Comparison of six D-dimer methods in patients suspected of deep vein thrombosis
    Gosselin, RC
    Owings, JT
    Kehoe, J
    Anderson, JT
    Dwyre, DM
    Jacoby, RC
    Utter, G
    Larkin, EC
    [J]. BLOOD COAGULATION & FIBRINOLYSIS, 2003, 14 (06) : 545 - 550
  • [7] D-dimer testing for deep venous thrombosis: A metaanalysis
    Heim, SW
    Schectman, JM
    Siadaty, MS
    Philbrick, JT
    [J]. CLINICAL CHEMISTRY, 2004, 50 (07) : 1136 - 1147
  • [8] Rapid D-Dimer assays to exclude deep venous thrombosis and pulmonary embolism:: Current status and new developments
    Janssen, MCH
    Wollersheim, H
    Verbruggen, B
    Nováková, IRO
    [J]. SEMINARS IN THROMBOSIS AND HEMOSTASIS, 1998, 24 (04) : 393 - 400
  • [9] KOOPMAN J, 1987, J LAB CLIN MED, V109, P75
  • [10] Clinically significant differences in the International Normalized Ratio measured with reagents of different sensitivities
    Lind, SE
    Pearce, LA
    Feinberg, WM
    Bovill, EG
    [J]. BLOOD COAGULATION & FIBRINOLYSIS, 1999, 10 (05) : 215 - 227