Comparative evaluation of the VERSANT HCV RNA 3.0, QUANTIPLEX HCV RNA 2.0, and COBAS AMPLICOR HCV MONITOR version 2.0 assays for quantification of hepatitis C virus RNA in serum

被引:30
作者
Germer, JJ
Heimgartner, PJ
Ilstrup, DM
Harmsen, WS
Jenkins, GD
Patel, R
机构
[1] Mayo Clin, Div Infect Dis, Rochester, MN 55905 USA
[2] Mayo Clin, Div Clin Microbiol, Rochester, MN USA
[3] Mayo Clin, Biostat Sect, Rochester, MN USA
关键词
D O I
10.1128/JCM.40.2.495-500.2002
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
A comparison of quantitative results expressed in hepatitis C virus (HCV) international units per milliliter, obtained from the VERSANT HCV RNA 3.0 (bDNA-3.0) assay, the QUANTIPLEX HCV RNA 2.0 (bDNA-2.0) assay, and the COBAS AMPLICOR HCV MONITOR version 2.0 (HCM-2.0) test was performed. A total of 168 patient specimens submitted to the Mayo Clinic Molecular Microbiology Laboratory for HCV quantification or HCV genotyping were studied. Of the specimens tested, 97, 88, and 79% yielded quantitative results within the dynamic range of the bDNA-3.0, bDNA-2.0, and HCM-2.0 assays, respectively. Overall, there was substantial agreement between the results generated by all three assays. A total of 15 out of 29 (52%) of the specimens determined to contain viral loads of <31,746 IU/ml by the bDNA-3.0 assay were categorized as containing viral loads within the range of 31,746 to 500,000 IU/ml by the bDNA-2.0 assay. Although substantial agreement was noted between the results generated by the bDNA-2.0 and bDNA-3.0 assays, a bias toward higher viral titer by the bDNA-2.0 assay was noted (P = 0.001). Likewise, although substantial agreement was noted between the results generated by the HCM-2.0 and bDNA-3.0 assays, a bias toward higher viral titer by the bDNA-3.0 assay was noted (P <=. 0.001). The discrepancy between the HCM-2.0 and bDNA-3.0 results was more pronounced when viral loads were >500,000 IU/ml and resulted in statistically significant differences (P less than or equal to 0.001) in determining whether viral loads were above or below 800,000 IU/ml of HCV RNA, the proposed threshold value for tailoring the duration of combination therapy. The expression of quantitative values in HCV international units per milliliter was a strength of both the bDNA-3.0 and HCM-2.0 assays.
引用
收藏
页码:495 / 500
页数:6
相关论文
共 19 条
[1]  
[Anonymous], 1999, J Hepatol, V30, P956
[2]   ISOLATION OF A CDNA CLONE DERIVED FROM A BLOOD-BORNE NON-A, NON-B VIRAL-HEPATITIS GENOME [J].
CHOO, QL ;
KUO, G ;
WEINER, AJ ;
OVERBY, LR ;
BRADLEY, DW ;
HOUGHTON, M .
SCIENCE, 1989, 244 (4902) :359-362
[3]   A COEFFICIENT OF AGREEMENT FOR NOMINAL SCALES [J].
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1960, 20 (01) :37-46
[4]   A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/ml [J].
Collins, ML ;
Irvine, B ;
Tyner, D ;
Fine, E ;
Zayati, C ;
Chang, CA ;
Horn, T ;
Ahle, D ;
Detmer, J ;
Shen, LP ;
Kolberg, J ;
Bushnell, S ;
Urdea, MS ;
Ho, DD .
NUCLEIC ACIDS RESEARCH, 1997, 25 (15) :2979-2984
[5]   Interferon alfa-2b alone or in combination with ribavirin for the treatment of relapse of chronic hepatitis C [J].
Davis, GL ;
Esteban-Mur, R ;
Rustgi, V ;
Hoefs, J ;
Gordon, SC ;
Trepo, C ;
Shiffman, ML ;
Zeuzem, S ;
Craxi, A ;
Ling, MH ;
Albrecht, J .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (21) :1493-1499
[6]   Accurate quantification of hepatitis C virus (HCV) RNA from all HCV genotypes by using branched-DNA technology [J].
Detmer, J ;
Lagier, R ;
Flynn, J ;
Zayati, C ;
Kolberg, J ;
Collins, M ;
Urdea, M ;
SanchezPescador, R .
JOURNAL OF CLINICAL MICROBIOLOGY, 1996, 34 (04) :901-907
[7]   Quantification of serum hepatitis C virus RNA [J].
Fang, JWS ;
Albrecht, JK ;
Jacobs, S ;
Lau, JYN .
HEPATOLOGY, 1999, 29 (03) :997-998
[8]   Determination of hepatitis C virus genotype by direct sequence analysis of products generated with the amplicor HCV test [J].
Germer, JJ ;
Rys, PN ;
Thorvilson, JN ;
Persing, DH .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (08) :2625-2630
[9]   Comparison of plasma virus loads among individuals infected with hepatitis C virus (HCV) genotypes 1, 2, and 3 by Quantiplex HCV RNA assay versions 1 and 2, Roche monitor assay, and an in-house limiting dilution method [J].
Hawkins, A ;
Davidson, F ;
Simmonds, P .
JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (01) :187-192
[10]   Improved version 2.0 qualitative and quantitative AMPLICOR reverse transcription-PCR tests for hepatitis C virus RNA: Calibration to international units, enhanced genotype reactivity, and performance characteristics [J].
Lee, SC ;
Antony, A ;
Lee, N ;
Leibow, J ;
Yang, JQ ;
Soviero, S ;
Gutekunst, K ;
Rosenstraus, M .
JOURNAL OF CLINICAL MICROBIOLOGY, 2000, 38 (11) :4171-4179