Scatter rejection by air gaps in diagnostic radiology. Calculations using a Monte Carlo collision density method and consideration of molecular interference in coherent scattering

被引:50
作者
Persliden, J
Carlsson, GA
机构
[1] Department of Radiation Physics, Faculty of Health Sciences, Linköping University
关键词
D O I
10.1088/0031-9155/42/1/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The air gap technique is an old method for scatter rejection. It is still used in lung examinations and may be reconsidered for use in digital radiography. Using magnification techniques, for example in mammography, the air gap thereby introduced simultaneously yields scatter rejection. A Monte Carlo collision density method is exploited to investigate the physical parameters relevant to this technique. Radiation quantities of scattered photons at points behind a water slab both on and laterally displaced from the central axis are calculated and their dependence on field area, slab thickness, air gap length and detector type is derived. Values of 'scatter-to-primary' ratios of the plane energy fluence (the energy imparted to a totally absorbing detector) are given for perpendicularly incident 30, 70 and 130 kV energy spectra, slab thicknesses of 0.05 and 0.2 m (30 kV: 0.05 m), air gaps of length 0.002-1.0 m and field areas from 8 x 10(-5) to 0.3 m(2). Contrast degradation factors are derived for both totally absorbing and thin detectors. The influence on the scatter-to-primary ratios of using divergent instead of parallel beams and of neglecting molecular interference in coherent scattering is analysed.
引用
收藏
页码:155 / 175
页数:21
相关论文
共 41 条
[1]  
[Anonymous], 1980, 33 ICRU
[2]   REFLECTION AND TRANSMISSION OF GAMMA RADIATION BY BARRIERS - MONTE-CARLO CALCULATION BY A COLLISION-DENSITY METHOD [J].
BERGER, MJ .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1955, 55 (06) :343-350
[3]  
BIRCH R, 1979, HOSPITAL PHYSICISTS, V30
[4]   QUANTITATIVE MEASUREMENT OF SMALL-ANGLE GAMMA-RAY SCATTERING FROM WATER, NYLON, AND LUCITE [J].
BRADLEY, DA ;
DANCE, DR ;
EVANS, SH ;
JONES, CH .
MEDICAL PHYSICS, 1989, 16 (06) :851-857
[5]   GENERALIZED USE OF CONTRAST DEGRADATION AND CONTRAST IMPROVEMENT FACTORS IN DIAGNOSTIC-RADIOLOGY - APPLICATION TO VANISHING CONTRAST [J].
CARLSSON, GA ;
CARLSSON, CA ;
NIELSEN, B ;
PERSLIDEN, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (07) :737-749
[6]   CALCULATION OF SCATTERING CROSS-SECTIONS FOR INCREASED ACCURACY IN DIAGNOSTIC-RADIOLOGY .1. ENERGY BROADENING OF COMPTON-SCATTERED PHOTONS [J].
CARLSSON, GA ;
CARLSSON, CA ;
BERGGREN, KF ;
RIBBERFORS, R .
MEDICAL PHYSICS, 1982, 9 (06) :868-879
[7]  
CARLSSON GA, ULIRADR049 LINK U
[8]   PERFORMANCE OF ANTISCATTER GRIDS IN DIAGNOSTIC-RADIOLOGY - EXPERIMENTAL MEASUREMENTS AND MONTE-CARLO SIMULATION STUDIES [J].
CHAN, HP ;
HIGASHIDA, Y ;
DOI, K .
MEDICAL PHYSICS, 1985, 12 (04) :449-454
[9]   CALCULATION OF DOSE AND CONTRAST FOR 2 MAMMOGRAPHIC GRIDS [J].
DANCE, DR ;
PERSLIDEN, J ;
CARLSSON, GA .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :235-248
[10]   X-RAY TRANSMISSION FORMULA FOR ANTISCATTER GRIDS [J].
DAY, GJ ;
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (12) :1429-1433