Postmortem analysis of bone growth into porous-coated acetabular components

被引:63
作者
Bloebaum, RD
Mihalopoulus, NL
Jensen, JW
Dorr, LD
机构
[1] Bone and Joint Research Laboratory, Vet. Administration Medical Center, Salt Lake City, UT
[2] Bone and Joint Research Laboratory, Vet. Administration Medical Center, Salt Lake City, UT 84148
[3] Univ. S. California Ctr. Arthr. J., Univ. of S. California Univ. Hosp., Health Care Consultation Center, Los Angeles, CA 90033-4634
关键词
D O I
10.2106/00004623-199707000-00007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Microradiography, backscattered electron microscopy, and histological analysis were used to conduct a quantitative postmortem study of seven consecutively retrieved anatomical porous replacement acetabular components that had been inserted during total hip arthroplasties. Screws had been used for the initial fixation of six components. The microradiographic analysis of all seven components showed that an average (and standard deviation) of 84 +/- 9 per cent (range, 72 to 93 per cent) of the porous coating was in direct apposition to the periprosthetic bone. The backscattered electron images demonstrated that an average of 12 +/- 6 per cent (range, 4 to 21 per cent) of the space available in the porous coating was occupied by ingrown bone. The amount of bone ingrowth was not significantly different among the three zones delineated by DeLee and Charnley. Uniformity of bone growth into the porous coating suggests that the preferential loading that occurs in the superior region did not differentially affect the bone ingrowth. The present study showed that consistent bone growth into anatomical porous replacement acetabular components can be achieved.
引用
收藏
页码:1013 / 1022
页数:10
相关论文
共 34 条
[1]  
BACHUS KN, 1992, CELL MATER, V2, P347
[2]   TOTAL KNEE ARTHROPLASTY IN THE MANAGEMENT OF PROXIMAL TIBIAL STRESS-FRACTURES [J].
TOMLINSON, MPW ;
DINGWALL, IM ;
PHILLIPS, H .
JOURNAL OF ARTHROPLASTY, 1995, 10 (05) :707-713
[3]  
BLOEBAUM RD, 1989, J HISTOTECHNOL, V12, P307
[4]   MINERAL APPOSITION RATES OF HUMAN CANCELLOUS BONE AT THE INTERFACE OF POROUS-COATED IMPLANTS [J].
BLOEBAUM, RD ;
BACHUS, KN ;
MOMBERGER, NG ;
HOFMANN, AA .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (05) :537-544
[5]  
BLOEBAUM RD, 1991, CLIN ORTHOP RELAT R, V268, P179
[6]  
BLOEBAUM RD, 1994, CLIN ORTHOPAEDICS, V309, P2
[7]  
BLOEBAUM RD, 1991, CLIN ORTHOP RELAT R, V267, P97
[8]   PERIACETABULAR STRESS DISTRIBUTIONS AFTER JOINT REPLACEMENT WITH SUBCHONDRAL BONE RETENTION [J].
CARTER, DR ;
VASU, R ;
HARRIS, WH .
ACTA ORTHOPAEDICA SCANDINAVICA, 1983, 54 (01) :29-35
[9]   MECHANICAL INFLUENCES IN BONE REMODELING - EXPERIMENTAL RESEARCH ON WOLFFS LAW [J].
CHAMAY, A ;
TSCHANTZ, P .
JOURNAL OF BIOMECHANICS, 1972, 5 (02) :173-&
[10]  
COOK SD, 1992, CLIN ORTHOP RELAT R, V283, P163