Comparison of I5M and 4Pi-microscopy

被引:88
作者
Bewersdorf, J. [1 ]
Schmidt, R. [1 ]
Hell, S. W. [1 ]
机构
[1] Max Planck Inst Biophys Chem, Dept NanoBiophoton, D-37077 Gottingen, Germany
关键词
confocal; deconvolution; fluorescence; interference; multiphoton; PSF; OTF; resolution; superresolution; two-photon; widefield;
D O I
10.1111/j.1365-2818.2006.01578.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
The axial (z-) resolution of similar to 100 nm provided by 4Pi and (IM)-M-5 fluorescence microscopy relies on the coherent addition of spherical wavefronts of two opposing high aperture angle lenses. Both microscopes feature a point-spread function (PSF) with a sharp central spot that is accompanied by axially shifted sidelobes which leads to replication artefacts in the raw image data. In a 4Pi-microscope the sidelobes are less pronounced than in (IM)-M-5 and without relevant lateral (x,y) substructure, making their posterior removal in the image reliable and fast. On the other hand, high speeds of raw data acquisition are more easily gained by (IM)-M-5. Moreover, (IM)-M-5 features a stronger signal as compared to the commonly employed two-photon excitation (2PE) 4Pi-imaging mode. We investigate here the capability of both techniques to image (aqueous) specimens without artefacts. To this end, we consider the optical transfer function (OTF) of the two microscopes in conjunction with the signal-to-noise-ratio (SNR) of the object to be imaged. The imaging of E. coli bacteria with an interconvertable setup enabled a direct comparison of the two imaging modes. As both systems rely on high aperture angles, water-immersion lenses of the largest numerical aperture available (NA = 1.2) were employed. The experimental results are corroborated by simulations assuming the signal strength encountered in the experiment. The comparison of the theoretical with the experimental PSFs/OTFs showed that our setup operated close to theory in both imaging modes. Although (IM)-M-5 provided about 10 times brighter raw image data as compared to (2PE) 4Pi-microscopy, the (IM)-M-5 data could not be entirely cleared of artefacts. In conclusion, with the current aperture angles and fluorescence signal strengths, it is not advisable to trade in the suppression of the sidelobes for a larger image signal.
引用
收藏
页码:105 / 117
页数:13
相关论文
共 33 条
[1]
Spatially modulated illumination microscopy allows axial distance resolution in the nanometer range [J].
Albrecht, B ;
Failla, AV ;
Schweitzer, A ;
Cremer, C .
APPLIED OPTICS, 2002, 41 (01) :80-87
[2]
ENHANCEMENT OF AXIAL RESOLUTION IN FLUORESCENCE MICROSCOPY BY STANDING-WAVE EXCITATION [J].
BAILEY, B ;
FARKAS, DL ;
TAYLOR, DL ;
LANNI, F .
NATURE, 1993, 366 (6450) :44-48
[3]
Bewersdorf J, 1998, J MICROSC-OXFORD, V191, P28
[4]
Axial superresolution with ultrahigh aperture lenses [J].
Blanca, CM ;
Hell, SW .
OPTICS EXPRESS, 2002, 10 (17) :893-898
[5]
2-PHOTON LASER SCANNING FLUORESCENCE MICROSCOPY [J].
DENK, W ;
STRICKLER, JH ;
WEBB, WW .
SCIENCE, 1990, 248 (4951) :73-76
[6]
Focal spots of size λ/23 open up far-field florescence microscopy at 33 nm axial resolution -: art. no. 163901 [J].
Dyba, M ;
Hell, SW .
PHYSICAL REVIEW LETTERS, 2002, 88 (16) :4-163901
[7]
Fluorescence microscopy with super-resolved optical sections [J].
Egner, A ;
Hell, SW .
TRENDS IN CELL BIOLOGY, 2005, 15 (04) :207-215
[8]
Fast 100-nm resolution three-dimensional microscope reveals structural plasticity of mitochondria in live yeast [J].
Egner, A ;
Jakobs, S ;
Hell, SW .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2002, 99 (06) :3370-3375
[9]
Development of a standing-wave fluorescence microscope with high nodal plane flatness [J].
Freimann, R ;
Pentz, S ;
Horler, H .
JOURNAL OF MICROSCOPY-OXFORD, 1997, 187 :193-200
[10]
Cooperative 4Pi excitation and detection yields sevenfold sharper optical sections in live-cell microscopy [J].
Gugel, H ;
Bewersdorf, J ;
Jakobs, S ;
Engelhardt, J ;
Storz, R ;
Hell, SW .
BIOPHYSICAL JOURNAL, 2004, 87 (06) :4146-4152