A multiple detector array helical x-ray microtomography system for specimen imaging

被引:10
作者
Corrigan, NM [1 ]
Chavez, AE [1 ]
Wisner, ER [1 ]
Boone, JM [1 ]
机构
[1] Univ Calif Davis, Dept Radiol, Xray Imaging Labs, Sacramento, CA 95817 USA
关键词
computed tomography; microtomography; mouse; helical CT; instrumentation;
D O I
10.1118/1.598662
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An x-ray computed microtomography system for specimen and small animal imaging was built and tested. The system used seventeen 48-mu m-wide detector arrays (a charge coupled device camera) and helical acquisition techniques. Images were acquired using 540 rays/view and 400 views/2 pi rr. The modulation transfer function (MTF) of the computed tomography images demonstrated 50 mu m limiting resolution, with MTF> 10% for objects larger than 60 mu m (similar to 8.3 cycles/mm). While soft tissue discrimination was compromised by a low signal-to-noise ratio, equine medullary bone core samples and the murine skeleton were well visualized. The incorporation of multiple detector arrays provided a 17-fold improvement in x-ray efficiency, which is a very important step toward improving the potential of microtomography as a scientific tool. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)03108-9].
引用
收藏
页码:1708 / 1713
页数:6
相关论文
共 20 条
[1]  
Beighley PE, 1997, ADV EXP MED BIOL, V430, P165
[2]   3D COMPUTED X-RAY TOMOGRAPHY OF HUMAN CANCELLOUS BONE AT 8 MU-M SPATIAL AND 10(-4) ENERGY RESOLUTION [J].
BONSE, U ;
BUSCH, F ;
GUNNEWIG, O ;
BECKMANN, F ;
PAHL, R ;
DELLING, G ;
HAHN, M ;
GRAEFF, W .
BONE AND MINERAL, 1994, 25 (01) :25-38
[3]   A FLUOROSCOPY-BASED COMPUTED-TOMOGRAPHY SCANNER FOR SMALL SPECIMEN RESEARCH [J].
BOONE, JM ;
ALEXANDER, GM ;
SEIBERT, JA .
INVESTIGATIVE RADIOLOGY, 1993, 28 (06) :539-544
[4]  
BOONE JM, 1996, MED PHYS, V23, P1084
[5]   Use of microcomputed tomography scanning as a new technique for the evaluation of membranous bone [J].
Buchman, SR ;
Sherick, DG ;
Goulet, RW ;
Goldstein, SA .
JOURNAL OF CRANIOFACIAL SURGERY, 1998, 9 (01) :48-54
[6]  
BUSHBERG JT, 1994, ESSENTIAL PHYSICS ME
[7]  
CORRIGAN NM, 1997, MED PHYS, V24, P1047
[8]  
Dainty JC., 1974, IMAGE SCI
[9]   ELASTICITY AND GEOMETRY MEASUREMENTS OF VASCULAR SPECIMENS USING A HIGH-RESOLUTION LABORATORY CT SCANNER [J].
DRANGOVA, M ;
HOLDSWORTH, DW ;
BOYD, CJ ;
DUNMORE, PJ ;
ROACH, MR ;
FENSTER, A .
PHYSIOLOGICAL MEASUREMENT, 1993, 14 (03) :277-290
[10]   HIGH SPATIAL-RESOLUTION IMAGING OF BONE-MINERAL USING COMPUTED MICROTOMOGRAPHY - COMPARISON WITH MICRORADIOGRAPHY AND UNDECALCIFIED HISTOLOGIC SECTIONS [J].
ENGELKE, K ;
GRAEFF, W ;
MEISS, L ;
HAHN, M ;
DELLING, G .
INVESTIGATIVE RADIOLOGY, 1993, 28 (04) :341-349