Small cantilevers for force spectroscopy of single molecules

被引:293
作者
Viani, MB [1 ]
Schäffer, TE
Chand, A
Rief, M
Gaub, HE
Hansma, PK
机构
[1] Univ Calif Santa Barbara, Dept Phys, Santa Barbara, CA 93106 USA
[2] Univ Munich, Lehrstuhl Angew Phys, D-80799 Munich, Germany
关键词
D O I
10.1063/1.371039
中图分类号
O59 [应用物理学];
学科分类号
摘要
We have used a simple process to fabricate small rectangular cantilevers out of silicon nitride. They have lengths of 9-50 mu m, widths of 3-5 mu m, and thicknesses of 86 and 102 nm. We have added metallic reflector pads to some of the cantilever ends to maximize reflectivity while minimizing sensitivity to temperature changes. We have characterized small cantilevers through their thermal spectra and show that they can measure smaller forces than larger cantilevers with the same spring constant because they have lower coefficients of viscous damping. Finally, we show that small cantilevers can be used for experiments requiring large measurement bandwidths, and have used them to unfold single titin molecules over an order of magnitude faster than previously reported with conventional cantilevers. (C) 1999 American Institute of Physics. [S0021-8979(99)04916-6].
引用
收藏
页码:2258 / 2262
页数:5
相关论文
共 26 条
[1]   MICROFABRICATION OF CANTILEVER STYLI FOR THE ATOMIC FORCE MICROSCOPE [J].
ALBRECHT, TR ;
AKAMINE, S ;
CARVER, TE ;
QUATE, CF .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1990, 8 (04) :3386-3396
[2]   ATOMIC FORCE MICROSCOPE [J].
BINNIG, G ;
QUATE, CF ;
GERBER, C .
PHYSICAL REVIEW LETTERS, 1986, 56 (09) :930-933
[3]   IRREVERSIBILITY AND GENERALIZED NOISE [J].
CALLEN, HB ;
WELTON, TA .
PHYSICAL REVIEW, 1951, 83 (01) :34-40
[4]  
CHAND A, UNPUB
[5]   Dynamic strength of molecular adhesion bonds [J].
Evans, E ;
Ritchie, K .
BIOPHYSICAL JOURNAL, 1997, 72 (04) :1541-1555
[6]   Fundamental noise limits for miniature acoustic and vibration sensors [J].
Gabrielson, TB .
JOURNAL OF VIBRATION AND ACOUSTICS-TRANSACTIONS OF THE ASME, 1995, 117 (04) :405-410
[7]   MECHANICAL-THERMAL NOISE IN MICROMACHINED ACOUSTIC AND VIBRATION SENSORS [J].
GABRIELSON, TB .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1993, 40 (05) :903-909
[8]   Thermal noise limitations on micromechanical experiments [J].
Gittes, F ;
Schmidt, CF .
EUROPEAN BIOPHYSICS JOURNAL WITH BIOPHYSICS LETTERS, 1998, 27 (01) :75-81
[9]   CALIBRATION OF ATOMIC-FORCE MICROSCOPE TIPS [J].
HUTTER, JL ;
BECHHOEFER, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (07) :1868-1873
[10]   HEIGHT CALIBRATION OF OPTICAL-LEVER ATOMIC-FORCE MICROSCOPES BY SIMPLE LASER INTERFEROMETRY [J].
JASCHKE, M ;
BUTT, HJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) :1258-1259