Suitability of LiF:Mg,Cu,P and Al2O3:C for low dose measurements in medical imaging

被引:14
作者
Duggan, L
Sathiakumar, C
Warren-Forward, H
Symonds, M
McConnell, P
Smith, T
Kron, T
机构
[1] Newcastle Mater Misericordiae Hosp, Dept Radiat Oncol, Hunter Reg Mail Ctr, Newcastle, NSW 2310, Australia
[2] St Vincents Hosp, Darlinghurst, NSW 2010, Australia
[3] Univ Newcastle, Dept Phys, Callaghan, NSW 2308, Australia
关键词
D O I
10.1093/oxfordjournals.rpd.a032888
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The aim of the study was to demonstrate the applicability of LiF:Mg,Cu,P and Al2O3:C in two low dose applications in medical imaging - neonate X ray examinations and dual-energy X ray absorptiometry (DXA). Dose response was investigated using multiple exposures and shown to be linear for both applications. Neonate phantom measurements were performed at 54 kV(p). Reproducibility was evaluated at 2.5 mA.s, a value characteristic of a neonatal chest X ray. In-phantom doses using exposure factors as low as 54 kV(p) and 1 mA.s (correlating to an approximate skin entrance dose of 35 +/- 10 mu Gy) can be assessed with reproducibility +/-20% (1 SD), the same order of magnitude as the reproducibility of the exposure. DXA was performed at 135 kV(p) on a Lunar Expert bone densitometer with a scan area of 18 cm x 17.3 cm. Skin entrance dose was calculated to be 440 mu Sv, by extrapolation from a calibration at higher doses on a Stabilipan II X ray therapy unit. Total body effective dose for an AP spine and femur scan were estimated as 56 mu Sv and 4 mu Sv/13 mu Sv (ovaries excluded/included), respectively.
引用
收藏
页码:425 / 428
页数:4
相关论文
共 21 条
[1]   Comparison of LiF:Mg,Cu,P (MCP-N, GR-200A) and alpha-Al2O3:C TL detectors in short-term measurements of natural radiation [J].
Budzanowski, M ;
Bilski, P ;
BotterJensen, L ;
Delgado, A ;
Olko, P ;
SaezVergara, JC ;
Waligorski, MPR .
RADIATION PROTECTION DOSIMETRY, 1996, 66 (1-4) :157-160
[2]  
HOROWITZ YS, 1993, RADIAT PROT DOSIM, V47, P135
[3]   Patient doses in bone mineral densitometry [J].
Huda, W ;
Morin, RL .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (821) :422-425
[4]  
International Commission on Radiological Protection, 1991, 60 ICRP
[5]  
International Commission on Radiological Protection, 1975, 23 ICRP
[6]  
JANSSEN RJJN, 1992, RADIAT PROT DOSIM, V43, P175
[7]  
Kron Tomas, 1993, Australasian Physical and Engineering Sciences in Medicine, V16, P137
[8]  
KUSUMA T, 1995, RADIAT PROT DOSIM, V58, P149
[9]  
McKeever S. W. S., 1995, Thermoluminescence Dosimetry Materials: Properties and Uses
[10]  
MCKINLAY AF, 1981, MED PHYSICS HDB, V5