Platelet Factor 4-Positive Thrombi Adhering to the Ventricles of a Ventricular Assist Device in Patients with Heparin-Induced Thrombocytopenia Type II

被引:3
作者
Beiras-Fernandez, A. [1 ]
Kanzler, I. [1 ]
Michel, S. [2 ]
Sadoni, S. [2 ]
Kilger, E. [2 ]
Beiras, A. [3 ]
Kur, F. [2 ]
机构
[1] JW Goethe Univ Hosp, Dept Cardiothorac & Vasc Surg, D-61590 Frankfurt, Germany
[2] Univ Munich, Dept Cardiac Surg, Munich, Germany
[3] Univ Santiago de Compostela, Dept Morphol Sci, Santiago, Spain
关键词
EXTRACORPOREAL MEMBRANE-OXYGENATION; ARGATROBAN; ANTIBODIES; MANAGEMENT; SUPPORT;
D O I
10.1016/j.transproceed.2013.01.045
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Background. Thromboembolism is a major complication in patients with ventricular assist devices (VADs). Drug anticoagulation and the use of biocompatible surfaces, such as coating with heparin, aim to reduce thromboembolism in these patients. Administration of heparin can lead to heparin-induced thrombocytopenia (HIT) type II, mainly through heparin/platelet factor 4 (PF4) antibodies. We assessed the presence of PF4 antibodies in VAD thrombi of patients with heparin-coated VADs and HIT II. Methods. Thrombi (n = 6) were obtained from the replaced Excor ventricles of patients with HIT II after biventricular.VAD implantation (Excor Adult; Berlin Heart, Germany). Excor ventricles were changed after clinical examination and suspicion of thrombi in the polyurethane valves. Expression of PF4- antibodies was assessed with the use of a polyclonal rabbit antibody (anti-PF4 antibody; Abcam, USA). Expression was assessed by 2 independent observers. Results. Biopsies of all thrombi showed an extreme positive immunoreaction for PF4. No differences between the different thrombi and localization (left/right Excor ventricle) were observed. The thrombi were organized, without lamination of fibrin and cellular layers. Conclusions. Platelet surface expression of PF4 in the thrombi reflects HIT antigen presentation. The physical relationship between the PF4-positive thrombi and the heparin-coated surface suggests that onset of HIT II could be influenced by the immobilized heparin coating.
引用
收藏
页码:2013 / 2016
页数:4
相关论文
共 13 条
[1]   Extracorporeal membrane oxygenation with danaparold sodium after massive pulmonary embolism [J].
Bauer, Christian ;
Vichova, Zuzana ;
Ffrench, Patrick ;
Hercule, Christiane ;
Jegaden, Olivier ;
Bastien, Olivier ;
Lehot, Jean-Jacques .
ANESTHESIA AND ANALGESIA, 2008, 106 (04) :1101-1103
[2]   Argatroban in extracorporeal membrane oxygenation [J].
Beiderlinden, Martin ;
Treschan, Tanja ;
Goerlinger, Klaus ;
Peters, Juergen .
ARTIFICIAL ORGANS, 2007, 31 (06) :461-465
[3]   Anticoagulative management of patients requiring left ventricular assist device implantation and suffering from heparin-induced thrombocytopenia type II [J].
Christiansen, S ;
Jahn, UR ;
Meyer, J ;
Scheld, HH ;
Van Aken, H ;
Kehrel, BE ;
Hammel, D .
ANNALS OF THORACIC SURGERY, 2000, 69 (03) :774-777
[4]   Lepirudin in heparin-induced thrombocytopenia and extracorporeal membranous oxygenation [J].
Dager, WE ;
Gosselin, RC ;
Yoshikawa, R ;
Owings, JT .
ANNALS OF PHARMACOTHERAPY, 2004, 38 (04) :598-601
[5]   Antibodies to Heparin-Platelet Factor 4 Complex: Pathogenesis, Epidemiology, and Management of Heparin-Induced Thrombocytopenia in Hemodialysis [J].
Davenport, Andrew .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2009, 54 (02) :361-374
[6]  
Dolch ME, 2010, EXP CLIN TRANSPLANT, V8, P329
[7]   The management of heparin-induced thrombocytopenia [J].
Keeling, D ;
Davidson, S ;
Watson, H .
BRITISH JOURNAL OF HAEMATOLOGY, 2006, 133 (03) :259-269
[8]   Prevalence and persistence of heparin/platelet factor 4 antibodies in patients with heparin coated and noncoated ventricular assist devices [J].
Koster, A ;
Sänger, S ;
Hansen, R ;
Sodian, R ;
Mertzlufft, F ;
Harke, C ;
Kuppe, H ;
Hetzer, R ;
Loebe, M .
ASAIO JOURNAL, 2000, 46 (03) :319-322
[9]   Heparin antibodies and thromboembolism in heparin-coated and noncoated ventricular assist devices [J].
Koster, A ;
Loebe, M ;
Sodian, R ;
Potapov, EV ;
Hansen, R ;
Müller, J ;
Mertzlufft, F ;
Crystal, GJ ;
Kuppe, H ;
Hetzer, R .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 121 (02) :331-335
[10]   Impact of heparin-induced thrombocytopenia on outcome in patients with ventricular assist device support: Single-institution experience in 358 consecutive patients [J].
Koster, Andreas ;
Huebler, Sabine ;
Potapov, Evgenij ;
Meyer, Oliver ;
Jurmann, Michael ;
Weng, Yuguo ;
Pasic, Miralem ;
Drews, Thorsten ;
Kuppe, Hermann ;
Loebe, Matthias ;
Hetzer, Roland .
ANNALS OF THORACIC SURGERY, 2007, 83 (01) :72-76