Abutment region dosimetry for the monoisocentric three-beam split field technique in the head and neck region using asymmetrical collimators

被引:8
作者
Abdel-Hakim, K
Nishimura, T
Takai, M
Suzuki, S
Sakahara, H
机构
[1] Hamamatsu Univ Sch Med, Dept Radiol, Hamamatsu, Shizuoka 4313192, Japan
[2] Hamamatsu Univ Sch Med, Dept Informat, Hamamatsu, Shizuoka 4313192, Japan
关键词
D O I
10.1259/bjr.75.893.750428
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Creating non-divergent field edges using asymmetric collimators and a single isocentre can improve matchline dosimetry owing to decreased reliance on operator skills and avoidance of couch movement. However, asymmetic jaws have an associated tolerance that can cause abutment to be misaligned. The matching area dose for monoisocentric three-beam split fields commonly used in head and neck cancer treatments using mismatched and matched collimators is the subject of this work. X-ray verification film was exposed in a solid-water phantom, and the dose at the matching area was evaluated using mismatched and matched collimators. In the case of mismatched (consistently overlapped) collimators, digital displays of an asymmetric collimator position within the tolerance indicated in the manufacturer's specifications were investigated for the three-beam split field technique. The effect of this technique on the junctional dose was also determined using matched collimators. Although the collimators showed a consistent overlap, a perfect dose distribution could be obtained at the matching area. The three-beam split field technique yielded an 8% overdose at the matchline using matched collimators. In conclusion, an awareness of the effects of the abutting technique and digital display tolerance is necessary to achieve good junction uniformity using asymmetric collimators.
引用
收藏
页码:428 / 434
页数:7
相关论文
共 15 条
[1]   The influence of beam energy on the outcome of postoperative radiotherapy in head and neck cancer patients: Secondary analysis of RTOG 85-03 [J].
Aref, A ;
Berkey, BA ;
Schwade, JG ;
Ensley, J ;
Schuller, DE ;
Haselow, RE ;
Ervin, TJ ;
Laramore, GE .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (02) :389-394
[2]   DOSE COMPUTATIONS FOR ASYMMETRIC FIELDS DEFINED BY INDEPENDENT JAWS [J].
CHUI, CS ;
MOHAN, R ;
FONTENLA, D .
MEDICAL PHYSICS, 1988, 15 (01) :92-95
[3]   Decreasing the dosimetric effects of misalignment when using a mono-isocentric technique for irradiation of head and neck cancer [J].
Fabrizio, PL ;
McCullough, EC ;
Foote, RL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (05) :1623-1634
[4]  
International Commission in Radiation Units and Measurements, 1993, 50 ICRU
[5]   DOSIMETRY OF ASYMMETRIC X-RAY COLLIMATORS [J].
KHAN, FM ;
GERBI, BJ ;
DEIBEL, FC .
MEDICAL PHYSICS, 1986, 13 (06) :936-941
[6]   COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40 [J].
KUTCHER, GJ ;
COIA, L ;
GILLIN, M ;
HANSON, WF ;
LEIBEL, S ;
MORTON, RJ ;
PALTA, JR ;
PURDY, JA ;
REINSTEIN, LE ;
SVENSSON, GK ;
WELLER, M ;
WINGFIELD, L .
MEDICAL PHYSICS, 1994, 21 (04) :581-618
[7]  
Lee P C, 1997, Med Dosim, V22, P59, DOI 10.1016/S0958-3947(96)00149-5
[8]  
LOSSHEK DD, 1988, MED PHYS, V15, P604
[9]  
LOSSHEK DD, 1988, MED PHYS, V15, P672
[10]   DOSIMETRIC CHARACTERISTICS OF A 6-MV PHOTON-BEAM FROM A LINEAR-ACCELERATOR WITH ASYMMETRIC COLLIMATOR JAWS [J].
PALTA, JR ;
AYYANGAR, KM ;
SUNTHARALINGAM, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (02) :383-387