Comparison of a mechanical acetabular alignment guide with computer placement of the socket

被引:244
作者
DiGioia, AM [1 ]
Jaramaz, B [1 ]
Plakseychuk, AY [1 ]
Moody, JE [1 ]
Nikou, C [1 ]
LaBarca, RS [1 ]
Levison, TJ [1 ]
Picard, F [1 ]
机构
[1] Western Penn Hosp, Inst Comp Assisted Orthopaed Surg, Pittsburgh, PA 15242 USA
关键词
alignment guides; acetabular component; total hip; surgical technique; computer-assisted surgery; navigation;
D O I
10.1054/arth.2002.30411
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We hypothesized that use of mechanical acetabular guides for intraoperative alignment leads to variations between the actual and desired implant orientation. Acetabular implant orientation using only the mechanical guide was studied in 78 patients (82 hills) undergoing primary total hip arthroplasty. A computer-assisted navigation system was used to measure alignment and to monitor the orientation of the pelvis during surgery. When using the mechanical guide, there was significant variation in cup alignment from the desired goal of 45degrees of abduction and 20degrees of flexion, and this would have resulted in unacceptable acetabular alignment in 78% of hips. With the support system used, there was significant variability in pelvic orientation during surgery. The mean anteversion of the pelvis was an average of 18degrees from the optimal orientation. These results show a clear need to develop more reliable tools than were used or anatomically based alignment strategies to provide reproducible and accurate acetabular alignment.
引用
收藏
页码:359 / 364
页数:6
相关论文
共 23 条
[1]   ANTEVERSION OF THE ACETABULAR CUP - MEASUREMENT OF ANGLE AFTER TOTAL HIP-REPLACEMENT [J].
ACKLAND, MK ;
BOURNE, WB ;
UHTHOFF, HK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (03) :409-413
[2]  
DIGIOIA AM, 1998, CLIN ORTHOP RELAT R, V354, P8
[3]  
Harkess WJ, 1992, CAMPBELLS ORTHOPAEDI, P54
[4]  
HARRIS WH, 1980, CLIN ORTHOP RELAT R, P188
[5]  
HERRLIN K, 1988, ACTA RADIOL, V29, P357
[6]  
HOIKKA V, 1993, ORTHOPEDICS, V16, P1321
[7]   Computer assisted measurement of cup placement in total hip replacement [J].
Jaramaz, B ;
DiGioia, AM ;
Blackwell, M ;
Nikou, C .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (354) :70-81
[8]  
Jaramaz B, 1999, LECT NOTES COMPUT SC, V1679, P876
[9]  
Jaramaz B, 1998, CLIN ORTHOP RELAT R, V355, P8
[10]   Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration [J].
Kennedy, JG ;
Rogers, WB ;
Soffe, KE ;
Sullivan, RJ ;
Griffen, DG ;
Sheehan, LJ .
JOURNAL OF ARTHROPLASTY, 1998, 13 (05) :530-534