A study of the wear resistance of three types of clinically applied UHMWPE for total replacement hip prostheses

被引:54
作者
Barbour, PSM [1 ]
Stone, MH [1 ]
Fisher, J [1 ]
机构
[1] Univ Leeds, Dept Mech Engn, Leeds LS2 9JT, W Yorkshire, England
关键词
wear; UHMWPE; total hip replacement; mechanical properties;
D O I
10.1016/S0142-9612(99)00096-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultra high molecular weight polyethylene (UHMWPE) is routinely used as one half of a bearing couple in clinical orthopaedic applications. This material is generally one of two grades, compression moulded GUR 1120, with a molecular weight of 4.4 x 10(6) g mol(-1), or ram extruded GUR 4150HP with a molecular weight of 7.3 x 10(6) g mol(-1) although other grades are used as well. This study examines the effect of molecular weight and processing method by comparing the wear resistance of these two materials together with a non-standard extruded form of GUR 1120. Wear studies have been conducted against counterface surfaces which represent new and scratch-damaged femoral heads with bovine serum lubricant in a uniaxial reciprocating test configuration. No difference in the wear resistance was found between any material when tested against smooth cobalt chrome counterfaces. On rough cobalt chrome counterfaces, although no statistically significant difference (P > 0.05) was found between the extruded and compression moulded GUR 1120, the extruded GUR 4150HP had a consistently better wear resistance. These results are discussed with reference to previous wear studies as well as the physical and mechanical properties of these materials. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:2101 / 2106
页数:6
相关论文
共 21 条
[1]  
BALDINI TH, 1997, T 43 ANN M ORTH RES, P780
[2]   THE INFLUENCE OF CONTACT STRESS ON THE WEAR OF UHMWPE FOR TOTAL REPLACEMENT HIP PROSTHESES [J].
BARBOUR, PSM ;
BARTON, DC ;
FISHER, J .
WEAR, 1995, 181 :250-257
[3]  
Berzen J., 1991, ULTRA HIGH MOL WEIGH, P50
[4]  
BROSSA F, 1996, T 5 WORLD BIOM C, V2, P197
[5]   LABORATORY WEAR TESTS AND CLINICAL OBSERVATIONS OF THE PENETRATION OF FEMORAL HEADS INTO ACETABULAR CUPS IN TOTAL REPLACEMENT HIP JOINTS .1. CHARNLEY PROSTHESES WITH POLYTETRAFLUOROETHYLENE ACETABULAR CUPS [J].
DOWSON, D ;
WALLBRIDGE, NC .
WEAR, 1985, 104 (03) :203-215
[6]   THE ROLE OF COUNTERFACE IMPERFECTIONS IN THE WEAR OF POLYETHYLENE [J].
DOWSON, D ;
TAHERI, S ;
WALLBRIDGE, NC .
WEAR, 1987, 119 (03) :277-293
[7]  
EVANS MA, 1996, T 5 WORLD BIOM C, V2, P186
[8]  
FARRAR DF, 1995, T 21 ANN M SOC BIOM, P232
[9]   PRELIMINARY-STUDY OF THE EFFECT OF AGING FOLLOWING IRRADIATION ON THE WEAR OF ULTRAHIGH-MOLECULAR-WEIGHT POLYETHYLENE [J].
FISHER, J ;
CHAN, KL ;
HAILEY, JL ;
SHAW, D ;
STONE, M .
JOURNAL OF ARTHROPLASTY, 1995, 10 (05) :689-692
[10]  
Fisher J, 1995, Proc Inst Mech Eng H, V209, P263, DOI 10.1243/PIME_PROC_1995_209_353_02