Feasibility of using the automatic generating system for quartz watches as a leadless pacemaker power source

被引:61
作者
Goto, H
Sugiura, T
Harada, Y
Kazui, T
机构
[1] Shizuoka Univ, Elect Res Inst, Hamamatsu, Shizuoka 4328011, Japan
[2] Hamamatsu Univ Sch Med, Hamamatsu, Shizuoka 4313124, Japan
关键词
accelerated motion; automatic power-generating system; button-type pacemaker; capacitor; charging characteristic; discharging characteristic; energy; kinetic energy; leadless pacemaker; mechano-electric energy conversion system;
D O I
10.1007/BF02513315
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
An automatic power-generating system (AGS) which converts kinetic energy into electric energy for quartz watches was tested as a power source for implantable cardiac pacemakers. An automatic power-generating mechanism and a capacitor (0.33 F) were removed from a quartz watch (SEIKO) and encapsulated in a polyvinyl case. Characteristics of the AGS were investigated by acceleration equipment. The capacitor in the AGS was charged to 2.0 V (0.66 J) by placing it on the equipment for about 30 minutes. The equipment has a 2 Hz cycle and generates +/-1.7 G at the end of each half cycle. The AGS (fully charged to 2.0 V) was used as the power source for a pulse generator circuit built using commercially available CMOS IC. The circuit generated pulses of 0.5 ms width at 7 Hz (60 pulses min(-1)). The voltage of the AGS was maintained at 1.6 V while it was being charged by the accelerations. The generator supplied pulses of 0.75 V, 1.47 mA through a 510 Omega load. With fully charged AGS, the generator was also used to pace a mongrel dog's heart at 140 beats min(-1) for 60 minutes. During pacing, the AGS supplied 420 mJ to the circuit and the cardiac muscle. The AGS was placed on the right ventricular wall of the mongrel dog under anaesthesia. Energy of 80 mJ is stored in a capacitor by the heart beating at about 200 beats/min for 30 minutes. Thus the AGS generated 13 mu J per heart beat. This result suggests that the AGS may supply enough energy for use in a cardiac pacemaker.
引用
收藏
页码:377 / 380
页数:4
相关论文
共 14 条
[1]  
ALT E, 1903, RATE ADAPTIVE CARDIA, P318
[2]  
Crossley GH, 1996, CIRCULATION, V94, P245
[3]   Lithium/carbon monofluoride (Li/CFx): A new pacemaker battery [J].
Greatbatch, W ;
Holmes, CF ;
Takeuchi, ES ;
Ebel, SJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (11) :1836-1840
[4]  
HAYAKAWA N, 1988, P C EUR CHRON, P81
[5]   High thresholds may alter end-of-life behavior in a dual chamber pacemaker [J].
Kraus, M ;
Buckingham, TA ;
Candinas, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (06) :1691-1697
[6]  
Lanmuller H, 1997, ARTIF ORGANS, V21, P210
[7]   Determinants of dual chamber pulse generators longevity [J].
Markewitz, A ;
Kronski, D ;
Kammeyer, A ;
Kaulbach, H ;
Weinhold, C ;
Doering, W ;
Reichart, B .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (12) :2116-2120
[8]  
MOND HG, 1992, PACE, V15, P95
[9]  
Preston T A, 1974, Med Instrum, V8, P322
[10]  
SCHALDACH M, 1975, ADV PACEMAKER TECHNO, pCH5