Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors

被引:105
作者
Gambarini, G [1 ]
机构
[1] Univ Roma La Sapienza, Dept Periodontol & Endodont, Rome, Italy
关键词
D O I
10.1097/00004770-200112000-00015
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
This study investigated the cyclic fatigue resistance of used nickel-titanium rotary instruments, clinically operated by a traditional high-torque motor and a new low-torque electric motor. Fifty instruments (ProFile, Maillefer, Baillagues, Switzerland) for each of the following sizes were used: .06-25, .06-20, and .04-25. They were divided into three groups: A = 10 new instruments (control), B = 20 used instruments operated by a high-torque motor, and C = 20 used instruments operated by a low-torque motor. Each instrument was used in 10 clinical cases (at least six molars). Cyclic fatigue testing of new and used instruments was performed with a device that allowed the instruments to rotate freely inside a stainless-steel artificial canal at 350 rpm. Data were recorded and statistically analyzed. Mean (and SD) numbers of rotation-to-breakage cycles for .06-25, .06-20, and .04-25 instruments were, respectively: group A 463 (85), 535 (101), and 613 (105); group B 229 (64), 343 (92), and 233 (84); and group C 352 (77), 454 (89), and 466 (86). Results showed that group A instruments exhibited significantly higher values of rotation-to-breakage cycles (two-sample t test, p < 0.01), when compared with groups B and C, respectively. Statistically higher resistance to cyclic fatigue was noted for group C in all sizes compared with group B (two-sample t test, p < 0.01), showing that the use of endodontic motor with lower torque values reduced cyclic fatigue of nickel-titanium rotary instruments.
引用
收藏
页码:772 / 774
页数:3
相关论文
共 11 条
[1]   TORSIONAL AND STIFFNESS PROPERTIES OF CANAL MASTER-U STAINLESS-STEEL AND NITINOL INSTRUMENTS [J].
CAMPS, J ;
PERTOT, WJ .
JOURNAL OF ENDODONTICS, 1994, 20 (08) :395-398
[2]  
Gambarini G, 2000, ENDOD DENT TRAUMATOL, V16, P95
[3]  
GAMBARINI G, 1999, J EVOLUT DENT ONLINE, V1, P4
[4]   Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments [J].
Haïkel, Y ;
Serfaty, R ;
Bateman, G ;
Senger, B ;
Allemann, C .
JOURNAL OF ENDODONTICS, 1999, 25 (06) :434-440
[5]  
MACHTOU P, 1997, CLINIC, V18, P253
[6]  
PONGIONE G, 1999, J DENT RES, V77, P397
[7]   Cyclic fatigue testing of nickel-titanium endodontic instruments [J].
Pruett, JP ;
Clement, DJ ;
Carnes, DL .
JOURNAL OF ENDODONTICS, 1997, 23 (02) :77-85
[8]  
SERENE TP, 1994, NICKEL TITANIUM INST
[9]   AN INITIAL INVESTIGATION OF THE BENDING AND TORSIONAL PROPERTIES OF NITINOL ROOT-CANAL FILES [J].
WALIA, H ;
BRANTLEY, WA ;
GERSTEIN, H .
JOURNAL OF ENDODONTICS, 1988, 14 (07) :346-351
[10]   Cyclic fatigue of Profile rotary instruments after simulated clinical use [J].
Yared, GM ;
Dagher, FEB ;
Machtou, P .
INTERNATIONAL ENDODONTIC JOURNAL, 1999, 32 (02) :115-119