Determination of drug concentrations in plasma by a highly automated, generic and flexible protein precipitation and liquid chromatography/tandem mass spectrometry method applicable to the drug discovery environment

被引:17
作者
O'Connor, D [1 ]
Clarke, DE [1 ]
Morrison, D [1 ]
Watt, AP [1 ]
机构
[1] Merck Sharp & Dohme Ltd, Res Labs, Dept Med Chem, Drug Metab Sect,Neurosci Res Ctr, Harlow CM20 2QR, Essex, England
关键词
D O I
10.1002/rcm.679
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This report presents a highly automated procedure for the determination of drug concentrations in plasma samples. The method is generic, in that it has been applied without adaptation to many different drug candidate molecules, but is also flexible, in that variations in the nature and number of samples to be analyzed can be readily accommodated. The method includes preparation of dilutions of analyte stock solutions, spiking these into control plasma to generate analytical standards, and preparation of samples suitable for analysis by high-performance liquid chromatography/tandem mass spectrometry (HPLC/MS/MS) by precipitation of plasma proteins with acetonitrile, centrifugation, and dilution of the supernatants with HPLC buffer. All of these steps, apart from centrifugation, are performed without manual intervention on an automated liquid-handling workstation using 96-well plates. Analysis is by HPLC/MS/MS, using a generic HPLC gradient. Commercially available software was used for optimization of parameters for analysis by HPLC/MS/MS, integration of chromatographic peaks, and quantification of drug concentrations. The use of this methodology in our laboratory has greatly facilitated the analysis of small sample sets for a large number of analytes, a situation regularly encountered in an early drug discovery environment. Copyright (C) 2002 John Wiley Sons, Ltd.
引用
收藏
页码:1065 / 1071
页数:7
相关论文
共 15 条
[1]  
Brockman A H, 2000, Curr Opin Drug Discov Devel, V3, P432
[2]  
Jemal M, 2000, BIOMED CHROMATOGR, V14, P422, DOI 10.1002/1099-0801(200010)14:6<422::AID-BMC25>3.0.CO
[3]  
2-I
[4]  
Jemal M, 1999, RAPID COMMUN MASS SP, V13, P97, DOI 10.1002/(SICI)1097-0231(19990130)13:2<97::AID-RCM461>3.0.CO
[5]  
2-T
[6]   Cassette-accelerated rapid rat screen: a systematic procedure for the dosing and liquid chromatography/atmospheric pressure ionization tandem mass spectrometric analysis of new chemical entities as part of new drug discovery [J].
Korfmacher, WA ;
Cox, KA ;
Ng, KJ ;
Veals, J ;
Hsieh, YS ;
Wainhaus, S ;
Broske, L ;
Prelusky, D ;
Nomeir, A ;
White, RE .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (05) :335-340
[7]   High-throughput high-performance liquid chromatography/mass spectrometry for modern drug discovery [J].
Kyranos, JN ;
Cai, H ;
Wei, D ;
Goetzinger, WK .
CURRENT OPINION IN BIOTECHNOLOGY, 2001, 12 (01) :105-111
[8]   Quantitative determination of L-775,606, a selective 5-hydroxytryptamine 1D agonist, in rat plasma using automated sample preparation and detection by liquid chromatography-tandem mass spectrometry [J].
Locker, KL ;
Morrison, D ;
Watt, AP .
JOURNAL OF CHROMATOGRAPHY B, 2001, 750 (01) :13-23
[9]   Matrix effect in quantitative LC/MS/MS analyses of biological fluids: A method for determination of finasteride in human plasma at picogram per milliliter concentrations [J].
Matuszewski, BK ;
Constanzer, ML ;
Chavez-Eng, CM .
ANALYTICAL CHEMISTRY, 1998, 70 (05) :882-889
[10]  
O'Connor D, 2002, CURR OPIN DRUG DI DE, V5, P52