Bead-based immunoassays with microelectrode detection

被引:42
作者
Farrell, S [1 ]
Ronkainen-Matsuno, NJ [1 ]
Halsall, HB [1 ]
Heineman, WR [1 ]
机构
[1] Univ Cincinnati, Dept Chem, Cincinnati, OH 45221 USA
关键词
microelectrode; detection; voltammetry; immunobeads; sandwich immunoassay;
D O I
10.1007/s00216-004-2632-z
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The suitability of a microelectrode as the detector for a small-volume, bead-based enzyme-labeled immunoassay for later use in a microfluidic device was investigated. The microelectrode helps to overcome consumption of the electroactive species by the electrode (depletion) that is encountered with macroelectrodes such as the rotating disk electrode (RDE) and allows the volume of the detection cell to be reduced. Microelectrodes also allow the chemical reactions to be monitored in real time due to the electrodes' close proximity to the assay site. A bead-based sandwich immunoassay for mouse IgG was developed with alkaline phosphatase (AP) as the enzyme label, p-aminophenyl phosphate (PAPP) as the enzyme substrate, and microelectrode detection. The diffusion coefficient of the product of enzymatic hydrolysis, p-aminophenol (PAP), was determined to be 7.2+/-0.9x10(-6) cm(2) s(-1). The detection limits were determined for free (0.52 ng mL(-1)) and bead-bound AP (10 ng mL(-1)). The number of binding sites for AP per bead was calculated to be 9.6x10(4) molecules/bead, and under saturation conditions the minimum detectable number of beads was 2500. Lower detection limits could be achieved with the microelectrode than the RDE while maintaining similar reproducibility. The microelectrode also made it possible to work with lower sample volumes (down to 10 muL) than with the RDE (minimum volume of 40 muL). Depletion of PAP was not observed with the microelectrode. The results obtained here with a microelectrode showed great promise for later use of microelectrodes in microfluidic devices with limited sample volumes. RDE detection cannot be used in a microfluidic system due to its complex set-up that includes a motor for rotation.
引用
收藏
页码:358 / 367
页数:10
相关论文
共 37 条
[1]   CHARACTERIZATION OF GLUCOSE MICROSENSORS FOR INTRACELLULAR MEASUREMENTS [J].
ABE, T ;
LAU, YY ;
EWING, AG .
ANALYTICAL CHEMISTRY, 1992, 64 (18) :2160-2163
[2]   Self-contained microelectrochemical immunoassay for small volumes using mouse IgG as a model system [J].
Aguilar, ZP ;
Vandaveer, WR ;
Fritsch, I .
ANALYTICAL CHEMISTRY, 2002, 74 (14) :3321-3329
[3]  
ASLAM M, 1998, BIOCONJUGATION
[4]  
Clark RA, 1998, ELECTROANAL CHEM, V20, P227
[5]  
CORNISHBOWDEN A, 1979, FUNDAMENTALS ENZYME
[6]   PULSE VOLTAMMETRY WITH MICROVOLTAMMETRIC ELECTRODES [J].
EWING, AG ;
DAYTON, MA ;
WIGHTMAN, RM .
ANALYTICAL CHEMISTRY, 1981, 53 (12) :1842-1847
[7]  
HALSALL HB, 1988, CLIN CHEM, V34, P1701
[8]   Oxidation state speciation of [ReI(DMPE)3]+/[ReII(DMPE)3]2+ by voltammetry with a chemically modified microelectrode [J].
Hu, ZM ;
Heineman, WR .
ANALYTICAL CHEMISTRY, 2000, 72 (11) :2395-2400
[9]   EXTENDING THE DETECTION LIMIT OF SOLID-PHASE ELECTROCHEMICAL ENZYME-IMMUNOASSAY TO THE ATTOMOLE LEVEL [J].
JENKINS, SH ;
HEINEMAN, WR ;
HALSALL, HB .
ANALYTICAL BIOCHEMISTRY, 1988, 168 (02) :292-299
[10]   AMPEROMETRIC MONITORING OF CHEMICAL SECRETIONS FROM INDIVIDUAL PANCREATIC BETA-CELLS [J].
KENNEDY, RT ;
HUANG, L ;
ATKINSON, MA ;
DUSH, P .
ANALYTICAL CHEMISTRY, 1993, 65 (14) :1882-1887