Evaluation of the ERETIC Method as an Improved Quantitative Reference for 1H HR-MAS Spectroscopy of Prostate Tissue

被引:54
作者
Albers, Mark J. [1 ,2 ]
Butler, Thomas N. [1 ]
Rahwa, Iman [1 ]
Bao, Nguyen [3 ]
Keshari, Kayvan R. [1 ]
Swanson, Mark G. [1 ]
Kurhanewicz, John [1 ,2 ]
机构
[1] Univ Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94158 USA
[2] Univ Calif Berkeley, Dept Bioengn, San Francisco, CA USA
[3] Varian Inc, Palo Alto, CA USA
基金
美国国家卫生研究院;
关键词
ERETIC; HR-MAS; TSP; prostate; MAGNETIC-RESONANCE SPECTROSCOPY; H-1-NMR SPECTROSCOPY; NMR ANALYSIS; STATE NMR; METABOLITES;
D O I
10.1002/mrm.21808
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Electronic REference To access In vivo Concentrations (ERETIC) method was applied to H-1 HR-MAS spectroscopy. The accuracy, precision, and stability of ERETIC as a quantitative reference were evaluated in solution and human prostate tissue samples. For comparison, the reliability of 3-(trimethylsilyl)propionic-2,2,3,3-d(4) acid (TSP) as a quantitation reference was also evaluated. The ERETIC and TSP peak areas were found to be stable in solution over the short-term and long-term, with long-term relative standard deviations (RSDs) of 4.10% and 2.60%, respectively. Quantification of TSP in solution using the ERETIC peak as a reference and a calibrated, rotor-dependent conversion factor yielded results with a precision <= 2.9% and an accuracy error <= 4.2% when compared with the expected values. The ERETIC peak area reproducibility was superior to TSP's reproducibility, corrected for mass, in both prostate surgical and biopsy samples (4.53% vs. 21.2% and 3.34% vs. 31.8%, respectively). Furthermore, the tissue TSP peaks exhibited only 27.5% of the expected area, which would cause an overestimation of metabolite concentrations if used as a reference. The improved quantification accuracy and precision provided by ERETIC may enable the detection of smaller metabolic differences that may exist between individual tissue samples and disease states. Magn Reson Med 61:525-532, 2009. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:525 / 532
页数:8
相关论文
共 18 条
[1]   Concentration measurement by proton NMR using the ERETIC method [J].
Akoka, S ;
Barantin, L ;
Trierweiler, M .
ANALYTICAL CHEMISTRY, 1999, 71 (13) :2554-2557
[2]   Improvement of the ERETIC method by digital synthesis of the signal and addition of a broadband antenna inside the NMR probe [J].
Akoka, S ;
Trierweiler, M .
INSTRUMENTATION SCIENCE & TECHNOLOGY, 2002, 30 (01) :21-29
[3]   Quantitative pathology in tissue MR spectroscopy based human prostate metabolomics [J].
Burns, MA ;
He, WL ;
Wu, CL ;
Cheng, LL .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2004, 3 (06) :591-598
[4]  
Cheng LL, 1998, CANCER RES, V58, P1825
[5]   Non-destructive quantitation of spermine in human prostate tissue samples using HRMAS 1H NMR spectroscopy at 9.4 T [J].
Cheng, LL ;
Wu, CL ;
Smith, MR ;
Gonzalez, RG .
FEBS LETTERS, 2001, 494 (1-2) :112-116
[6]   High-resolution magic angle spinning 1H NMR spectroscopy and reverse transcription-PCR analysis of apoptosis in a rat glioma [J].
Griffin, JL ;
Blenkiron, C ;
Valonen, PK ;
Caldas, C ;
Kauppinen, RA .
ANALYTICAL CHEMISTRY, 2006, 78 (05) :1546-1552
[7]   Generating long-lasting 1H and 13C hyperpolarization in small molecules with parahydrogen-induced polarization [J].
Jonischkeit, Thorsten ;
Bommerich, Ute ;
Stadler, Joerg ;
Woelk, Klaus ;
Niessen, Heiko G. ;
Bargon, Joachim .
JOURNAL OF CHEMICAL PHYSICS, 2006, 124 (20)
[8]   QUANTITATION OF METABOLITES IN HUMAN BLOOD-SERUM BY PROTON MAGNETIC-RESONANCE SPECTROSCOPY - A COMPARATIVE-STUDY OF THE USE OF FORMATE AND TSP AS CONCENTRATION STANDARDS [J].
KRIAT, M ;
CONFORTGOUNY, S ;
VIONDURY, J ;
SCIAKY, M ;
VIOUT, P ;
COZZONE, PJ .
NMR IN BIOMEDICINE, 1992, 5 (04) :179-184
[9]  
Millis K, 1999, MAGNET RESON MED, V41, P257, DOI 10.1002/(SICI)1522-2594(199902)41:2<257::AID-MRM8>3.0.CO
[10]  
2-N