Characterisation of an antibody coated microcantilever as a potential immuno-based biosensor

被引:96
作者
Grogan, C [1 ]
Raiteri, R
O'Connor, GM
Glynn, TJ
Cunningham, V
Kane, M
Charlton, M
Leech, D
机构
[1] Natl Univ Ireland Univ Coll Galway, Dept Phys, Galway, Ireland
[2] Univ Genoa, Dept Biophys & Elect Engn, I-16145 Genoa, Italy
[3] Natl Univ Ireland Univ Coll Galway, Natl Diagnost Ctr, Galway, Ireland
[4] Natl Univ Ireland Univ Coll Galway, Dept Chem, Galway, Ireland
关键词
biosensor characterisation; microcantilever; antibody immobilisation; enzyme linked immunosorbent assay; fluorescence microscope;
D O I
10.1016/S0956-5663(01)00276-7
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In this study, we investigated the activity, stability, lifetime and re-usability of monoclonal antibodies to myoglobin covalently immobilised onto microfabricated cantilever surfaces. These sensing surfaces are of interest to us in the development of novel cantilever-based immunosensors. For such sensors the antibody layer represents the sensing element while the microcantilever acts as a mechanical transducer. A procedure for producing re-usable biological coatings has been tested with different independent techniques. An Enzyme Linked Immunosorbent Assay (ELISA) was used to determine the presence of an active antibody coating, and to monitor the lifetime and stability of the immobilised antibody. Through this analysis, the activity of the immobilised antibody layer was found to be more stable with the introduction of sucrose, as a stabilising agent. Sucrose was applied to the immobilised antibody layer after each regeneration step. The immobilised antibody was found to have a stable active lifetime for up to 7 weeks. Fluorescence microscopy was used to give information on the distribution of the coating on the gold and silicon nitride sides of the cantilever. Atomic Force Microscopy was used to determine the presence of the biological coating on the cantilever and to obtain information on the surface morphology of the biological element of the sensor. The combined results provide valuable information on the development of an optimised sensing element and demonstrate a set of methods to use for future sensor-to-sensor characterisation. Preliminary experimental results showing the antibody activity against myoglobin, detected with a microcantilever based sensor prototype confirmed the motivations and potentialities of the proposed immunosensing technique. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:201 / 207
页数:7
相关论文
共 17 条
  • [1] A biosensor based on micromechanical interrogation of living cells
    Antonik, MD
    DCosta, NP
    Hoh, JH
    [J]. IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1997, 16 (02): : 66 - 72
  • [2] Micromechanical thermogravimetry
    Berger, R
    Lang, HP
    Gerber, C
    Gimzewski, JK
    Fabian, JH
    Scandella, L
    Meyer, E
    Guntherodt, HJ
    [J]. CHEMICAL PHYSICS LETTERS, 1998, 294 (4-5) : 363 - 369
  • [3] Surface stress in the self-assembly of alkanethiols on gold
    Berger, R
    Delamarche, E
    Lang, HP
    Gerber, C
    Gimzewski, JK
    Meyer, E
    Guntherodt, HJ
    [J]. SCIENCE, 1997, 276 (5321) : 2021 - 2024
  • [4] A sensitive method to measure changes in the surface stress of solids
    Butt, HJ
    [J]. JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1996, 180 (01) : 251 - 260
  • [5] Translating biomolecular recognition into nanomechanics
    Fritz, J
    Baller, MK
    Lang, HP
    Rothuizen, H
    Vettiger, P
    Meyer, E
    Güntherodt, HJ
    Gerber, C
    Gimzewski, JK
    [J]. SCIENCE, 2000, 288 (5464) : 316 - 318
  • [6] Mechanical resonant immunospecific biological detector
    Ilic, B
    Czaplewski, D
    Craighead, HG
    Neuzil, P
    Campagnolo, C
    Batt, C
    [J]. APPLIED PHYSICS LETTERS, 2000, 77 (03) : 450 - 452
  • [7] JONSSON U, 1985, BIOCHEM J, V227, P363
  • [8] Microcantilever-based biosensors
    Moulin, AM
    O'Shea, SJ
    Welland, ME
    [J]. ULTRAMICROSCOPY, 2000, 82 (1-4) : 23 - 31
  • [9] Measuring surface-induced conformational changes in proteins
    Moulin, AM
    O'Shea, SJ
    Badley, RA
    Doyle, P
    Welland, ME
    [J]. LANGMUIR, 1999, 15 (26) : 8776 - 8779
  • [10] NICOLINI C, 1992, PHYSICS WORLD MAY