Line selection and interference correction for the analysis of tungsten alloy by inductively coupled plasma atomic emission spectrometry

被引:4
作者
Bae, ZU [1 ]
Lee, SH [1 ]
Lee, SH [1 ]
机构
[1] KOREA TUNGSTEN CO,CENT RES INST,DALSUNG 711860,KYUNGPOOK,SOUTH KOREA
关键词
inductively coupled plasma atomic emission spectrometry; interference correction; line selection; tungsten alloy;
D O I
10.1016/S0039-9140(96)02007-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method for the analysis of tungsten alloy to determine selected elements using inductively coupled plasma atomic emission spectroscopy is described, with emphasis on line selection and spectral interference. The spectral interference coefficients were calculated for the spectral lines of selected major and trace elements. These values were used to select analytical lines and to calibrate concentrtions of the analytes, The detection limits of the elements for this method were determined and compared with those obtained by flame atomic absorption spectrometry and direct current carbon are emission spectrometry. The results indicated that the detection limits for all of the elements determined by the proposed method are significantly better than those obtained by other techniques. In this study, the analytical reliability of the proposed method was estimated by comparison of the analytical data for the two types of tungsten alloys produced by the Korean Tungsten Company with those obtained by the matrix matching method and the results indicated that the accuracy of multi-element analysis is satisfactory.
引用
收藏
页码:47 / 51
页数:5
相关论文
共 25 条
[1]   DETERMINATION OF CADMIUM IN BIOLOGICAL SAMPLES BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY AFTER EXTRACTION WITH 1,5-BIS(DI-2-PYRIDYLMETHYLENE) THIOCARBONOHYDRAZIDE [J].
ALMENDRO, JME ;
OJEDA, CB ;
DETORRES, AG ;
PAVON, JMC .
ANALYST, 1992, 117 (11) :1749-1751
[2]  
Bever M. B., 1986, ENCY MATERIALS ENG, P4159
[4]  
BOTTO RI, 1981, DEV ATOMIC PLASMA SP, P141
[5]  
BOUMANS PWJ, 1987, INDUCTIVELY COUPLE 1, P421
[6]   DIRECT TRACE-ELEMENT ANALYSIS OF TUNGSTEN POWDERS, ALLOYS AND RELATED MATERIALS BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY (ICP-AES) [J].
BRENNER, IB ;
ERLICH, S ;
VIAL, G ;
MCCORMACK, J ;
GROSDAILLON, P ;
ASHER, AE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (06) :637-644
[7]   LINE SELECTION AND DETERMINATION OF TRACE AMOUNTS OF ELEMENTS IN TUNGSTEN BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
CARRE, M ;
DERODRIGUEZ, OD ;
MERMET, JM ;
BRIDENNE, M ;
MAROT, Y .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (01) :49-55
[8]  
Danninger H., 1986, International Journal of Refractory & Hard Metals, V5, P144
[9]   EVALUATION OF INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY AS AN ELEMENT-SPECIFIC DETECTOR FOR SUPERCRITICAL FLUID CHROMATOGRAPHY [J].
FORBES, KA ;
VECCHIARELLI, JF ;
UDEN, PC ;
BARNES, RM .
ANALYTICAL CHEMISTRY, 1990, 62 (18) :2033-2037
[10]  
Guanqyao Z., 1992, ANAL LETT, V25, P561, DOI [10.1080/00032719208016115, DOI 10.1080/00032719208016115]