Automated continuous monitoring of drug dissolution process using a flow injection on-line dialysis sampling - solvent extraction separation spectrophotometric system

被引:3
作者
Fang, Q
Sun, YQ
Fang, ZL
机构
[1] Northeastern Univ, Res Ctr Analyt Sci, Shenyang 110006, Peoples R China
[2] Shenyang Pharmaceut Univ, Shenyang 110015, Peoples R China
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1999年 / 364卷 / 04期
关键词
Ethambutol; Bromocresol; Chinese Pharmacopoeia; Microdialysis Sampling; Bromocresol Green;
D O I
10.1007/s002160051347
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection (FI) system was developed incorporating on-line solvent extraction separation and stopped-flow dialysis systems for automated continuous monitoring of multi-vessel drug dissolution processes. The system employed PTFE pump tubes for solvent delivery, and combined a coaxial phase segmentor, conical cavity gravitational phase separator with a microdialysis sampling system to produce a reliable extraction system capable of stable non-stop operation over extended period of time. The system was evaluated for continuous monitoring in multi-vessel drug dissolution testing of ethambutol hydrochloride tablets using bromocresol green as reagent and chloroform as extractant. With the six-vessel dissolution system used, a sampling frequency of 15/h for each dissolution vessel was achieved with a total of 90 determinations per hour, giving an average precision of 3.6% RSD (n = 11). Results showed good agreement with those obtained using the Chinese Pharmacopoeia standard method.