Variability of Global Left Ventricular Deformation Analysis Using Vendor Dependent and Independent Two-Dimensional Speckle-Tracking Software in Adults

被引:186
作者
Risum, Niels [1 ]
Ali, Sophia [1 ]
Olsen, Niels T. [2 ]
Jons, Christian [2 ]
Khouri, Michel G. [1 ]
Lauridsen, Trine K. [1 ]
Samad, Zainab [1 ]
Velazquez, Eric J. [1 ]
Sogaard, Peter [2 ]
Kisslo, Joseph [1 ]
机构
[1] Duke Univ, Med Ctr, Div Cardiovasc Dis, Durham, NC 27705 USA
[2] Gentofte Univ Hosp, Dept Cardiol, Gentofte, Denmark
关键词
Strain; Speckle-tracking; Validation; Reproducibility; CARDIAC RESYNCHRONIZATION THERAPY; MYOCARDIAL STRAIN-MEASUREMENT; HYPERTROPHIC CARDIOMYOPATHY; ECHOCARDIOGRAPHY; DYSSYNCHRONY; VALIDATION; INFARCTION;
D O I
10.1016/j.echo.2012.08.007
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Background: Evaluation of myocardial deformation by two-dimensional speckle-tracking is useful for clinical and research purposes. However, differences may exist among different ultrasound machines, software packages, frame rates, and observers. Methods: Thirty patients underwent echocardiography on both GE (Vivid 9; GE Vingmed Ultrasound AS, Horten, Norway) and Philips (iE33; Philips Ultrasound, Bothell, WA) ultrasound systems. From each study, two sets of images were stored in Digital Imaging and Communications in Medicine format, optimized for strain evaluation: one set of images at the acquisition frame rate (55-90 frames/sec) and one set of images at a compressed frame rate of 30 frames/sec. Vendor-independent software (VIS; TomTec 2D Cardiac Performance Analysis, Munich, Germany) was used to measure strain in multiple directions and was compared with vendor-specific software (GE EchoPAC; GE Vingmed Ultrasound AS). Results: Intraobserver and interobserver coefficients of variation ranged from 5.5% to 8.7% for longitudinal strain, from 10.7% to 20.8% for circumferential, and from 15.3% to 33.4% for radial and transverse strain. Strain values obtained using VIS were comparable with those obtained using vendor-specific software for longitudinal strain, regardless of ultrasound machine or frame rate. For circumferential strain, a consistent large bias was observed between VIS and vendor-specific software, with higher values using VIS. Slightly higher strain values were observed by analysis at the acquisition frame rate compared with the low frame rate, but no consistent bias was observed between images from different vendors. Conclusions: Global longitudinal strain consistently showed good reproducibility, while reproducibility was moderate for circumferential strain and poor in the radial direction. Retrospective analysis of legacy Digital Imaging and Communications in Medicine data at 30 frames/sec can be reliably performed for longitudinal strain. (J Am Soc Echocardiogr 2012;25:1195-203.)
引用
收藏
页码:1195 / 1203
页数:9
相关论文
共 23 条
[1]
Noninvasive myocardial strain measurement by speckle tracking echocardiography - Validation against sonomicrometry and tagged magnetic resonance imaging [J].
Amundsen, BH ;
Helle-Valle, T ;
Edvardsen, T ;
Torp, H ;
Crosby, J ;
Lyseggen, E ;
Stoylen, A ;
Ihlen, H ;
Lima, JAC ;
Smiseth, OA ;
Slordahl, SA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (04) :789-793
[2]
Feasibility and Accuracy of Different Techniques of Two-Dimensional Speckle Based Strain and Validation With Harmonic Phase Magnetic Resonance Imaging [J].
Bansal, Manish ;
Cho, Goo-Yeong ;
Chan, Jonathan ;
Leano, Rodel ;
Haluska, Brian A. ;
Marwick, Thomas H. .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2008, 21 (12) :1318-1325
[3]
Analysis of myocardial deformation based on ultrasonic pixel tracking to determine transmurality in chronic myocardial infarction [J].
Becker, Michael ;
Hoffmann, Rainer ;
Kuehl, Harald P. ;
Grawe, Helena ;
Katoh, Markus ;
Kramann, Rafael ;
Buecker, Arno ;
Hanrath, Peter ;
Heussen, Nicole .
EUROPEAN HEART JOURNAL, 2006, 27 (21) :2560-2566
[4]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]
Left ventricular strain and strain rate: characterization of the effect of load in human subjects [J].
Burns, Andrew T. ;
La Gerche, Andre ;
D'hooge, Jan ;
MacIsaac, Andrew I. ;
Prior, David L. .
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2010, 11 (03) :283-289
[6]
Global 2-Dimensional Strain as a New Prognosticator in Patients With Heart Failure [J].
Cho, Goo-Yeong ;
Marwick, Thomas H. ;
Kim, Hyun-Sook ;
Kim, Min-Kyu ;
Hong, Kyung-Soon ;
Oh, Dong-Jin .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (07) :618-624
[7]
D'hooge J, 2000, Eur J Echocardiogr, V1, P154, DOI 10.1053/euje.2000.0031
[8]
Strain analysis in patients with severe aortic stenosis and preserved left ventricular ejection fraction undergoing surgical valve replacement [J].
Delgado, Victoria ;
Tops, Laurens F. ;
van Bommel, Rutger J. ;
van der Kley, Frank ;
Marsan, Nina Ajmone ;
Klautz, Robert J. ;
Versteegh, Michel I. M. ;
Holman, Eduard R. ;
Schalij, Martin J. ;
Bax, Jeroen J. .
EUROPEAN HEART JOURNAL, 2009, 30 (24) :3037-3047
[9]
Impaired Myocardial Radial Function in Asymptomatic Patients with Type 2 Diabetes Mellitus: A Speckle-Tracking Imaging Study [J].
Ernande, Laura ;
Rietzschel, Ernst R. ;
Bergerot, Cyrille ;
De Buyzere, Marc L. ;
Schnell, Frederic ;
Groisne, Laure ;
Ovize, Michel ;
Croisille, Pierre ;
Moulin, Philippe ;
Gillebert, Thierry C. ;
Derumeaux, Genevieve .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2010, 23 (12) :1266-1272
[10]
Reproducibility and Inter-Vendor Variability of Left Ventricular Deformation Measurements by Three-Dimensional Speckle-Tracking Echocardiography [J].
Gayat, Etienne ;
Ahmad, Homaa ;
Weinert, Lynn ;
Lang, Roberto M. ;
Mor-Avi, Victor .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2011, 24 (08) :878-885