On-line computerized vectorcardiography: Influence of body position, heart rate, radiographic contrast fluid and myocardial ischemia

被引:23
作者
Jensen, SM
Haggmark, S
Johansson, G
Naslund, U
机构
[1] UMEA UNIV HOSP, DEPT THORAC SURG, S-90185 UMEA, SWEDEN
[2] UMEA UNIV HOSP, DEPT ANESTHESIOL, S-90185 UMEA, SWEDEN
关键词
myocardial ischemia; monitoring; vectorcardiography; atrial pacing; contrast fluid; body position;
D O I
10.1159/000177422
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
On-line computerized vectorcardiography (cVCG) is increasingly being used for continuous monitoring of myocardial ischemia, however, little is known about factors other than ischemia causing electrocardiographic abnormalities. This paper describes how three important cVCG parameters, STC-VM, ST-VM and QRS-VD, are affected by different body positions, myocardial ischemia, contrast injection and increasing heart rate in patients with and without coronary artery disease. The main findings of the study are: contrast injection and different body positions caused major changes in QRS-VT but affected ST-VM and STC-VM to a minor degree. Increasing heart rate by atrial pacing produced substantial changes in all three parameters. Ischemia during angioplasty also produced changes in all three parameters, STC-VM being the most sensitive parameter. In conclusion: (1) STC-VM (greater than or equal to 50 mu V) is the most valuable parameter for monitoring ischemia; (2) we propose ST-VM greater than or equal to 50 mu V as criterion instead of previously used 25 mu V; (3) QRS-VD cannot be used as a single marker of ischemia, and (4) electrocardiographic changes induced by increased heart rate should be taken into account during interpretation.
引用
收藏
页码:576 / 584
页数:9
相关论文
共 29 条
[1]   DETECTION OF TRANSIENT MYOCARDIAL-ISCHEMIA BY COMPUTER-ANALYSIS OF STANDARD AND SIGNAL-AVERAGED HIGH-FREQUENCY ELECTROCARDIOGRAMS IN PATIENTS UNDERGOING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
ABBOUD, S ;
COHEN, RJ ;
SELWYN, A ;
GANZ, P ;
SADEH, D ;
FRIEDMAN, PL .
CIRCULATION, 1987, 76 (03) :585-596
[2]  
BESJAKOV J, 1993, CARDIAC EFFECTS ENRI
[4]  
CRIBIER A, 1983, CLIN CARDIOL, V6, P17
[5]   CORRELATION BETWEEN CHANGES IN R-WAVE AMPLITUDE AND LEFT-VENTRICULAR VOLUME INDUCED BY RAPID ATRIAL-PACING [J].
DANIELS, S ;
ISKANDRIAN, AS ;
HAKKI, AH ;
KANE, SA ;
BEMIS, CE ;
HOROWITZ, LN ;
GREENSPAN, AM ;
SEGAL, BL .
AMERICAN HEART JOURNAL, 1984, 107 (04) :711-717
[6]  
DELLBORG M, 1991, CORONARY ARTERY DIS, V2, P43
[7]   DYNAMIC QRS AND ST-SEGMENT CHANGES IN MYOCARDIAL-INFARCTION MONITORED BY CONTINUOUS ONLINE VECTORCARDIOGRAPHY [J].
DELLBORG, M ;
RIHA, M ;
SWEDBERG, K .
JOURNAL OF ELECTROCARDIOLOGY, 1991, 23 :11-19
[8]   DYNAMIC CHANGES OF THE QRS COMPLEX IN UNSTABLE ANGINA-PECTORIS [J].
DELLBORG, M ;
GUSTAFSSON, G ;
RIHA, M ;
SWEDBERG, K .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1992, 36 (02) :151-162
[9]   AN ACCURATE, CLINICALLY PRACTICAL SYSTEM FOR SPATIAL VECTORCARDIOGRAPHY [J].
FRANK, E .
CIRCULATION, 1956, 13 (05) :737-749
[10]  
GRENDAHL H, 1972, ACTA MED SCAND, V191, P493