A 3-D SAR model for current source interstitial hyperthermia

被引:38
作者
deBree, J [1 ]
VanderKoijk, JF [1 ]
Lagendijk, JJW [1 ]
机构
[1] UNIV UTRECHT HOSP,DEPT RADIOTHERAPY,NL-3508 GA UTRECHT,NETHERLANDS
关键词
D O I
10.1109/10.536905
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A three-dimensional (3-D) model is presented for the calculation of the specific absorption rate (SAR) in human tissue during current source interstitial hyperthermia. The model is capable of millimeter resolution and can cope with irregular implants in heterogeneous tissue. The SAR distribution is calculated from the electrical potential, The potential distribution is determined by the dielectric properties of the tissue and by the electrode configuration, The dielectric properties and the current injection of the electrodes are represented on a 3-D uniform grid, The calculated potential at an electrode current injection point is not the actual electrode potential at that point. To estimate this potential a grid independent representation of an electrode together with an analytical solution in the neighborhood of the electrode are used, The calculated potential on the electrode surface is used to estimate the electrode impedance, The tissue implementation is validated by comparing calculated distributions with analytical solutions, The electrode implementation is verified by comparing different discretizations of an electrode configuration and by comparing numerically calculated electrode impedances with analytically calculated impedances.
引用
收藏
页码:1038 / 1045
页数:8
相关论文
共 25 条
[1]   NUMERICAL-ANALYSIS OF THE MAGNETIC-FIELD FOR ARBITRARY MAGNETIC-SUSCEPTIBILITY DISTRIBUTIONS IN 3D [J].
BHAGWANDIEN, R ;
MOERLAND, MA ;
BAKKER, CJG ;
BEERSMA, R ;
LAGENDIJK, JJW .
MAGNETIC RESONANCE IMAGING, 1994, 12 (01) :101-107
[2]  
CORRY PM, 1990, BRACHYTHERAPY HDR LD
[3]   TEMPERATURE UNIFORMITY DURING HYPERTHERMIA - THE IMPACT OF LARGE VESSELS [J].
CREZEE, J ;
LAGENDIJK, JJW .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) :1321-1337
[4]  
Croft D.R., 1977, HEAT TRANSFER CALCUL
[5]  
DEBREE J, IN PRESS BR J RADIOL
[6]   APPLICATION OF A CAPACITIVE-COUPLING INTERSTITIAL HYPERTHERMIA SYSTEM AT 27 MHZ - STUDY OF DIFFERENT APPLICATOR CONFIGURATIONS [J].
DEURLOO, IKK ;
VISSER, AG ;
MORAWSKA, M ;
VANGEEL, CAJF ;
VANRHOON, GC ;
LEVENDAG, PC .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (01) :119-132
[7]  
DIRKS HK, 1988, KAPAZITATSKOEFFICIEN
[8]  
Doss J D, 1976, Med Instrum, V10, P16
[9]  
DUFORT EC, 1953, MATH TABLES OTHER AI, V7, P135, DOI DOI 10.2307/2002754
[10]  
HAND JW, 1991, HYPERTHERMIA ONCOLOG, V3