Ambulatory monitoring of the impedance cardiogram

被引:175
作者
Willemsen, GHM [1 ]
DeGeus, EJC [1 ]
Klaver, CHAM [1 ]
VanDoornen, LJP [1 ]
Carroll, D [1 ]
机构
[1] VRIJE UNIV AMSTERDAM, DEPT PSYCHON, AMSTERDAM, NETHERLANDS
关键词
ambulatory monitoring; impedance cardiography; systolic time intervals; stroke volume; cardiac output; heart rate; motility;
D O I
10.1111/j.1469-8986.1996.tb02122.x
中图分类号
B84 [心理学];
学科分类号
04 ; 0402 ;
摘要
The growing need for more advanced ambulatory monitoring has led to the development of an ambulatory monitor for impedance cardiography (VU-AMD). This paper presents two studies addressing the validity of the VU-AMD. In the first study, the cardiovascular responses of 25 subjects during various conditions were simultaneously recorded with the VU-AMD and a standard laboratory impedance device. Correlations between the responses of the ambulatory and laboratory devices were high, both inter- and intraindividually, except for stroke volume and cardiac output during exercise. In the second study, 26 subjects underwent 24-hr monitoring with the VU-AMD. The values obtained with the VU-AMD were realistic and varied in a predictable way over activity and posture. It is concluded that the VU-AMD is a valid device for the measurement of systolic time intervals in real-life situations, but its applicability for absolute stroke volume and cardiac output determination remains to be established.
引用
收藏
页码:184 / 193
页数:10
相关论文
共 30 条
[1]  
[Anonymous], INDIVIDUAL DIFFERENC
[2]   COMPARISON OF SPOT AND BAND IMPEDANCE CARDIOGRAM ELECTRODES ACROSS DIFFERENT TASKS [J].
BOOMSMA, DI ;
DEVRIES, J ;
ORLEBEKE, JF .
PSYCHOPHYSIOLOGY, 1989, 26 (06) :695-699
[3]  
DEGEUS EJC, 1990, PSYCHOPHYSIOLOGY, V27, P457
[4]  
DOERR BM, 1981, AVIAT SPACE ENVIR MD, V52, P394
[5]  
FAHRENBERG J, 1993, PSYCHOPHYSIOLOGY, V7, P275
[6]   EFFECTS OF ADRENERGIC RECEPTOR ACTIVATION AND BLOCKADE ON SYSTOLIC PREEJECTION PERIOD HEART RATE AND ARTERIAL PRESSURE IN MAN [J].
HARRIS, WS ;
SCHOENFELD, CD ;
WEISSLER, AM .
JOURNAL OF CLINICAL INVESTIGATION, 1967, 46 (11) :1704-+
[7]   SIGNAL FIDELITY REQUIREMENTS FOR DERIVING IMPEDANCE CARDIOGRAPHIC MEASURES OF CARDIAC-FUNCTION OVER A BROAD HEART-RATE RANGE [J].
HURWITZ, BE ;
SHYU, LY ;
LU, CC ;
REDDY, SP ;
SCHNEIDERMAN, N ;
NAGEL, JH .
BIOLOGICAL PSYCHOLOGY, 1993, 36 (1-2) :3-21
[8]  
JOHNSON DW, 1993, PSYCHOPHYSIOLOGY, V7, P209
[9]  
KAMATH MV, 1991, MED SCI SPORT EXER, V23, P428
[10]  
KUBICEK WG, 1974, BIOMED ENG, V9, P410