A HPMT based set-up to characterize scintillating crystals

被引:12
作者
D'Ambrosio, C [1 ]
Ercoli, C
Jaaskelainen, S
Lecoeur, G
Leutz, H
Loos, R
Piedigrossi, D
Puertolas, D
Rosso, E
Schomaker, R
机构
[1] CERN, Div EP, CH-1211 Geneva 23, Switzerland
[2] Ist Nazl Fis Nucl, Rome, Italy
[3] DEP, Roden, Netherlands
关键词
D O I
10.1016/S0168-9002(99)00539-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have developed a fully automatic measurement set-up, capable of measuring light yields arising from scintillating crystals in a linear range of about four orders of magnitude. The photodetector is a hybrid photomultiplier tube specially developed to optimize linear range and photon detection. Crystal and photodetector are temperature controlled by a closed water circuit, as this is essential when measuring low light yield scintillating crystals with a marked temperature dependence of their light yield. Gamma sources can be placed either on top or on the side of the crystal. In this latter case, the source can be automatically moved by a computer-controlled step motor to provide a uniformity profile of the light yield along the crystal. Tagged and not-tagged operation modes are possible. The whole set-up is computer-controlled in an effort to provide fast and reliable measurements, to characterize many crystals per day. This is important for the quality control of the lead tungstate crystals that will be applied in the electromagnetic calorimeter of the CMS-detector at the LHC at CERN. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:387 / 398
页数:12
相关论文
共 14 条
[1]  
AMBROSIO CD, 1995, ECALPWOLY01 CERN
[2]  
*ATLAS COLL, 1994, 9443 CERNLHCC ATLAS
[3]   Precise determination of the light yield of scintillating crystals [J].
Baccaro, S ;
Barone, LM ;
Borgia, B ;
Castelli, F ;
Cavallari, F ;
deNotaristefani, F ;
Diemoz, M ;
Faccini, R ;
Festinesi, A ;
Leonardi, E ;
Longo, E ;
Montecchi, M ;
Organtini, G ;
Pacciani, L ;
Pirro, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 385 (01) :69-73
[4]  
*CMS COLL, 1994, 9438 CERNLHCC CMS CO
[5]  
*CMS COLL, 1997, 9733 CERNLHCC CMS CO
[6]   The ISPA-tube and the HPMT, two examples of a new class of photodetectors: the Hybrid Photo Detectors. [J].
D'Ambrosio, C ;
Gys, T ;
Heijne, E ;
Leutz, H ;
Piedigrossi, D ;
Puertolas, D ;
Rosso, E .
NUCLEAR PHYSICS B, 1998, :638-643
[7]  
DAMBROSIO C, 1994, NUCL INSTRUM METH A, V338, P389, DOI 10.1016/0168-9002(94)91321-8
[8]   FURTHER RESULTS ON PHOTOELECTRON COUNTING WITH SMALL-DIAMETER SCINTILLATING FIBERS [J].
DAMBROSIO, C ;
GYS, T ;
LEUTZ, H ;
PIEDIGROSSI, D ;
PUERTOLAS, D ;
TAILHARDAT, S ;
SHIMIZU, T ;
SHINJI, O .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 345 (02) :279-283
[9]   Low dose-rate irradiation set-up for scintillating crystals [J].
DAmbrosio, C ;
Ercoli, C ;
Jaaskelainen, S ;
Rosso, E ;
Wicht, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 388 (1-2) :119-126
[10]   BACKSCATTERING OF 10-100 KEV ELECTRONS FROM THICK TARGETS [J].
DARLINGTON, EH .