Reproducible fabrication of miniaturized glucose sensors: preparation of sensing membranes for continuous monitoring

被引:32
作者
Chung, TD
Jeong, RA
Kang, SK
Kim, HC
机构
[1] Seoul Natl Univ, Coll Med, Dept Biomed Engn, Chongno Gu, Seoul 110799, South Korea
[2] Seoul Natl Univ, Med Res Ctr, Inst Med & Biol Engn, Chongno Gu, Seoul 110799, South Korea
关键词
glucose oxidase; poly(1,3-diaminobenzene); EQCM; poly-L-lysine; PTFE;
D O I
10.1016/S0956-5663(01)00185-3
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The immobilization process of glucose oxidase(GOx) in the poly(1,3-diaminobenzene) (poly(1,3-DAB)) network was closely investigated in situ using an electrochemical quartz crystal microbalance(EQCM). GOx captured in approximately 50 nm thick poly-1,3-DAB layer causes a 514 Hz frequency increase, corresponding to 541 ng, and distributes mostly in the outer part of the polymer film. The presence Of poly-L-lysine and glutaraldehyde during electropolymerization of poly(1,3-DAB) improves sensitivity by raising the amount of GOx immobilized. Adding a protective membrane on to the enzyme layer from poly(tetrafluroethylene) (PTFE) dispersed in aqueous media lets the entire fabrication procedure finish perfectly without nonaqueous solvent. The finalized needle-type glucose sensors show competent functions in sensitiviy, stability, biocompatibility, lifetime, interference and reproducibility. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1079 / 1087
页数:9
相关论文
共 21 条
[1]   TRANSIENT-RESPONSE OF MULTILAYER ELECTROENZYMATIC BIOSENSORS [J].
BACHA, S ;
BERGEL, A ;
COMTAT, M .
ANALYTICAL CHEMISTRY, 1995, 67 (10) :1669-1678
[2]   DESIGN AND INVITRO STUDIES OF A NEEDLE-TYPE GLUCOSE SENSOR FOR SUBCUTANEOUS MONITORING [J].
BINDRA, DS ;
ZHANG, YN ;
WILSON, GS ;
STERNBERG, R ;
THEVENOT, DR ;
MOATTI, D ;
REACH, G .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1692-1696
[3]   Electropolymerization of hydroxybenzene and aminobenzene isomers on platinum electrodes to assemble interference-free electrochemical biosensors [J].
Carelli, I ;
Chiarotto, I ;
Curulli, A ;
Palleschi, G .
ELECTROCHIMICA ACTA, 1996, 41 (11-12) :1793-1800
[4]   Electropolymerization of amphiphilic monomers for designing amperometric biosensors [J].
Cosnier, S .
ELECTROANALYSIS, 1997, 9 (12) :894-902
[5]   Biomolecule immobilization on electrode surfaces by entrapment or attachment to electrochemically polymerized films. A review [J].
Cosnier, S .
BIOSENSORS & BIOELECTRONICS, 1999, 14 (05) :443-456
[6]   USE OF POLYMER-FILMS IN AMPEROMETRIC BIOSENSORS [J].
EMR, SA ;
YACYNYCH, AM .
ELECTROANALYSIS, 1995, 7 (10) :913-923
[7]   ELECTROPOLYMERIZED FILMS TO PREVENT INTERFERENCES AND ELECTRODE FOULING IN BIOSENSORS [J].
GEISE, RJ ;
ADAMS, JM ;
BARONE, NJ ;
YACYNYCH, AM .
BIOSENSORS & BIOELECTRONICS, 1991, 6 (02) :151-160
[8]  
Imisides MD, 1996, CHEMTECH, V26, P19
[9]  
Iwuoha EI, 1999, ELECTROANAL, V11, P367, DOI 10.1002/(SICI)1521-4109(199905)11:5<367::AID-ELAN367>3.3.CO
[10]  
2-S