Experimental limitations in impedance spectroscopy .5. Apparatus contributions and corrections

被引:52
作者
Edwards, DD
Hwang, JH
Ford, SJ
Mason, TO
机构
[1] NORTHWESTERN UNIV,DEPT MAT SCI & ENGN,EVANSTON,IL 60208
[2] NORTHWESTERN UNIV,MAT RES CTR,EVANSTON,IL 60208
[3] NORTHWESTERN UNIV,CTR ADV CEMENT BASED MAT,EVANSTON,IL 60208
关键词
impedance spectroscopy; apparatus; simulation;
D O I
10.1016/S0167-2738(97)00206-3
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Overlooked apparatus (cabling, leads, sample holder) contributions to experimental impedance spectra can be significant, especially at high frequencies, and can obscure the true sample response. Instrumental Limitations are discussed and high-frequency artifacts arising from apparatus contributions are investigated as they pertain to accurate impedance measurements of materials systems. Remediation strategies are presented, including geometrical adjustments, lead shielding, and null-correction procedures.
引用
收藏
页码:85 / 93
页数:9
相关论文
共 16 条
  • [1] STUDY OF SOLID ELECTROLYTE POLARIZATION BY A COMPLEX ADMITTANCE METHOD
    BAUERLE, JE
    [J]. JOURNAL OF PHYSICS AND CHEMISTRY OF SOLIDS, 1969, 30 (12) : 2657 - &
  • [2] BOUKAMP BA, 1990, EQUIVCRT DEP CHEM TE
  • [3] IMPEDANCE SPECTROSCOPY OF HYDRATING CEMENT-EASED MATERIALS - MEASUREMENT, INTERPRETATION, AND APPLICATION
    CHRISTENSEN, BJ
    COVERDALE, RT
    OLSON, RA
    FORD, SJ
    GARBOCZI, EJ
    JENNINGS, HM
    MASON, TO
    [J]. JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 1994, 77 (11) : 2789 - 2804
  • [4] COOPER EA, 1992, THESIS NW U EVANSTON
  • [5] Experimental limitations in impedance spectroscopy .3. Effect of reference electrode geometry/position
    Hsieh, G
    Mason, TO
    Garboczi, EJ
    Pederson, LR
    [J]. SOLID STATE IONICS, 1997, 96 (3-4) : 153 - 172
  • [6] Experimental limitations in impedance spectroscopy .2. Electrode artifacts in three-point measurements on Pt/YSZ
    Hsieh, G
    Mason, TO
    Pederson, LR
    [J]. SOLID STATE IONICS, 1996, 91 (3-4) : 203 - 212
  • [7] Experimental limitations in impedance spectroscopy .1. Simulation of reference electrode artifacts in three-point measurements
    Hsieh, G
    Ford, SJ
    Mason, TO
    Pederson, LR
    [J]. SOLID STATE IONICS, 1996, 91 (3-4) : 191 - 201
  • [8] Experimental limitations in impedance spectroscopy .4. Electrode contact effects
    Hwang, JH
    Kirkpatrick, KS
    Mason, TO
    Garboczi, EJ
    [J]. SOLID STATE IONICS, 1997, 98 (1-2) : 93 - 104
  • [9] Characterization of humidity-sensing NiO-Ni(OH)(2) nanocomposites by impedance spectroscopy
    Hwang, JH
    Mason, TO
    Buehler, MF
    Darab, JG
    Matson, DW
    Linehan, JC
    [J]. MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 1995, 204 (1-2): : 252 - 257
  • [10] Irvine J. T. S., 1990, Advanced Materials, V2, P132, DOI 10.1002/adma.19900020304